Repetition-level data, clinician annotations and analysis code behind the paper
Pavlikov, A.; Petrosyan, V.; Agapov, V.; Gorodnichev, M.; Lobunko, D.; Skvortsov, D. Markerless On-Device Detection of Compensatory Movement Patterns in Upper-Limb Rehabilitation Exercises from Monocular RGB Video: A Validation Study in Healthy Adults. Sensors 2026, 26(16), 5054. doi:10.3390/s26165054
The paper evaluates a browser-based pipeline that estimates body pose from a single RGB camera and flags five compensatory movement patterns during a seated lateral arm-raise exercise. This repository holds everything needed to recompute the reported results, and a script that checks each number against the value printed in the paper.
Every headline figure in the manuscript is reproduced by:
python3 code/reproduce.pyNo dependencies beyond the Python standard library; Python 3.8 or newer. The
run takes about ten seconds and ends with either All values match the manuscript. or a list of the values that did not.
data/
repetitions.csv 913 repetitions: system output plus the per-repetition
quantities it is derived from (621 detection, 292 calibration)
annotations.csv 1242 labels: two clinicians x 621 repetitions
consensus.csv which repetitions the clinicians fully agreed on
participants.csv cohort membership and repetition counts
thresholds.csv the calibrated operating points and the timing rule
DATA_DICTIONARY.md every column, its units and its meaning
study-instrument/
OVERVIEW.md what the instrument is, and what is deliberately absent
webapp/ the recording and annotation application used to collect the data
analysis.py detection metrics and the threshold sweep
protocol.md the recording protocol as executed
code/
reproduce.py recomputes the reported numbers and verifies them
figure_elbow.py the elbow criterion behind Figure 4 of the paper
figure_spread.py how unevenly the detector performs
figure_confusion.py agreement with each clinician, class by class
extract_repetition_tables.py
how data/ was derived from the raw recordings
figures/
elbow_threshold.* the rendered figures, vector (PDF) and raster (PNG)
spread.*
confusion.*
docs/
protocol.md recording protocol
annotation_guide.md what the clinicians were asked to do
privacy.md what is published, what is withheld, and why
The video recordings. They are identifiable video of participants. The consent obtained covers research use and publication of anonymised results; it does not cover public release of the recordings themselves. They remain on a restricted laboratory server under pseudonymous codes.
The deployed application. What is published under study-instrument/ is the research
build that produced these data — the recording and annotation software, and the offline
recomputation that defines every metric. Its live configuration is withheld because it held
the access tokens and given names of the two annotating clinicians; config.example.json
carries placeholders in its place.
The per-frame keypoint series. Derived skeletal coordinates are health data and, as continuous per-person kinematics, are not equivalent to the aggregated results the consent contemplates. They are available from the corresponding author on request, for research use.
This is why the tables are at the repetition level. That granularity is not a
compromise on reproducibility: the per-repetition sustained values published
here are a sufficient statistic for the 250 ms decision rule, so the detector's
output can be recovered at any threshold. reproduce.py verifies that identity
against the published flags on all 621 detection repetitions, with no
disagreements. See "Sustained values" in the data dictionary.
The inertial recordings behind the angular-validation study, which was carried out earlier and separately, are no longer retained. The agreement statistics in the paper are those obtained in that analysis and cannot be recomputed here. The paper states this and treats the affected analyses as limitations rather than results.
Twenty-seven healthy adults, in three non-overlapping cohorts: 9 for threshold calibration, 18 for detection validation, and 10 in the earlier angular study. Seated, each participant performed a lateral arm raise under six instructed conditions — once correctly, and once for each of the five compensations. A clinician instructed and physically demonstrated each condition; condition order was randomised; co-occurring compensations were permitted rather than suppressed.
Two clinicians then annotated every repetition independently, blind to the instructed condition, and marked every compensation they could see, not just the instructed one. That is why a repetition instructed to be compensation-free may carry a label, and why a flag raised outside the instructed condition is not automatically a false positive.
Thresholds were fixed on the 9-participant calibration cohort by Youden's index subject to a false-positive rate no higher than 10%, then frozen before the 18-participant detection cohort was scored. No threshold was tuned on the data it was evaluated on.
Scored against each clinician over all 621 repetitions — the primary analysis:
| Compensation | F1 vs R1 | F1 vs R2 | Maturity as stated in the paper |
|---|---|---|---|
| Trunk lean | 0.92 | 0.92 | near-expert |
| Head tilt | 0.90 | 0.87 | near-expert |
| Incomplete elbow extension | 0.72 | 0.73 | moderate |
| Inter-limb asymmetry | 0.66 | 0.60 | moderate |
| Shoulder elevation | 0.55 | 0.50 | research-grade, not recommended for deployment |
| macro-F1 | 0.75 | 0.72 |
Scoring instead on the 477 repetitions the clinicians fully agreed on gives macro-F1 0.78. That is the favourable reading, and the paper reports it as an upper bound rather than as the result.
The macro-average deliberately mixes signs of different maturity and should not be read as a single deployable-performance figure. Per-participant macro-F1 ranges from 0.38 to 0.96 (median 0.75).
reproduce.py additionally verifies the per-class confusion matrices against
both clinicians, the shoulder-elevation false-positive rate, the elbow
criterion behind Figure 4 (median minimum elbow angle 142° on elbow-flexion
repetitions against 166° on those instructed correct, area under the ROC curve
0.92, sensitivity 0.70 at specificity 0.95 at the operating point), the
calibration cohort's discrimination, and the participant-clustered bootstrap
intervals.
Figure 4 of the paper, rendered from data/repetitions.csv alone. The left
panel shows the minimum elbow angle reached in each repetition, per
participant, split by instructed condition; the right panel the ROC curve with
the calibrated operating point marked. The overlap either side of the dashed
line is the point: the threshold sits on a narrow distribution in healthy
participants, which is what makes its placement — not the metric's
discrimination — the fragile part.
The same result, disaggregated. On the left, one macro-F1 per participant: the median is 0.75, but individual participants run from 0.38 to 0.96, so a single headline number describes no particular person well. On the right, one F1 per sign against each clinician, which is the evidence behind the paper's three tiers — trunk lean and head tilt near-expert, elbow extension and inter-limb asymmetry moderate, shoulder elevation research-grade and not recommended for deployment. Averaging across signs of such different maturity is exactly what the paper warns against reading as a deployable-performance figure.
Full counts behind every F1 above, shaded within the clinician's row so the shading reads as recall and specificity rather than being swamped by the true negatives. Trunk lean misses one positive out of 89 against either clinician; shoulder elevation produces 65 and 81 false positives, which is the failure the paper attributes to a structural confound with normal abduction rather than to threshold placement.
The rendered files are committed under figures/ in both PDF and PNG, so
nothing needs to be run to see them. To rebuild:
python3 -m pip install matplotlib
python3 code/figure_elbow.py
python3 code/figure_spread.py
python3 code/figure_confusion.pyEach script draws from data/ alone and imports its scoring from
reproduce.py, so no figure can drift away from the verified numbers.
code/extract_repetition_tables.py is the script that produced data/ from
the raw archive. It is published so the derivation can be audited, though it
cannot be run without the raw keypoint recordings. It carries the full geometry:
the isotropic coordinate convention, the resting-baseline definition, the gating
rules, and the 250 ms accumulation.
One detail worth flagging, since it is visible in the output. Coordinates from
the pose estimator are normalised per axis (x = px/W, y = px/H), which on a
16:9 frame scales the axes differently and does not preserve in-plane angles.
All geometry here therefore rescales x by W/H first. This correction was
made during peer review, and every number in this repository is post-correction.
Participation was documented on the institutional informed-consent form for research participation of the Federal Center of Brain Research and Neurotechnologies (FMBA of Russia), approved by order No. 117 of 1 June 2020. Personal data, including video as biometric personal data, were processed under Federal Law No. 152-FZ of 27 July 2006. Under the applicable institutional requirements, studies of this type in healthy volunteers without medical intervention do not require separate ethics-committee review. Participants gave separate consent for video recording, for processing of video as biometric personal data, and for use of anonymised results in publications, and retain the right to withdraw consent and request deletion.
Identifiers in this repository are pseudonymous codes; the key linking them to
individuals is held separately by the responsible investigator and is not part
of this release. See docs/privacy.md.
- Healthy volunteers simulating compensations on instruction. Reduced range of motion, spasticity, pathological synergies and variable speed are absent, and the strongest metrics are the large-amplitude gestures that are easiest to exaggerate — so their agreement is the most likely to fall on real, subtler compensations.
- A single exercise. Segmentation, thresholds and metric quality do not transfer to upper-limb rehabilitation in general.
- One camera configuration. Height, distance, viewing angle and lighting were not varied, so device independence is not established empirically.
- The segmentation state machine was never validated against an independent reference, so repetition-count accuracy and phase-boundary error are unknown.
- The aggregate quality score reported in the paper is an unvalidated, equally-weighted summary and is not included here as an outcome.
- Anthropometry and handedness were not recorded, so the claim that the metrics limit dependence on body size follows from their construction and is not tested on these data.
Code in code/ is released under the MIT Licence (LICENSE).
Data in data/ are released under CC BY 4.0 (LICENSE-DATA).
Attribution is required for the data. If you use them, cite the paper.
Please cite the paper. The archive itself carries two DOIs:
10.5281/zenodo.21747009— all versions; resolves to the latest. Cite this one unless you need to pin an exact version.10.5281/zenodo.21747010— version 1.0.0 specifically.
Author and affiliation metadata is in CITATION.cff.
Corresponding author: a.e.pavlikov@mtuci.ru
Requests for the per-frame keypoint recordings, for research use, go to the same address.