23 changes: 23 additions & 0 deletions ipApp/Db/XGS600-gauge.template
Original file line number Diff line number Diff line change
@@ -0,0 +1,23 @@
#
# Database to read gauge pressure from Agilent XGS600 controllers.
#
# Parameters:
# $(SYSTEM): PV prefix
# $(CONTROLLER): Controller name from st.cmd
# $(CHANNEL): Channel number
#

record(ai, "$(SYSTEM):getPressure")
{
field(DESC, "Reads Cell $(CONTROLLER) Pressure")
field(SCAN, "1 second")
field(DTYP, "stream")
field(INP, "@XGS600.proto getPressure($(CHANNEL)) $(CONTROLLER)")
field(PINI, "YES")
field(EGU, "mbar")
field(VAL, "0")
field(PREC, "3")
field(LOLO, "1.333E-12") field(LLSV, "MAJOR")
field(HIGH, "1E-7") field(HSV, "MINOR")
field(HIHI, "1E-6") field(HHSV, "MAJOR")
}
96 changes: 96 additions & 0 deletions ipApp/Db/XGS600-setpoints.template
Original file line number Diff line number Diff line change
@@ -0,0 +1,96 @@
#
# Agilent XGS600 controllers setpoints control system.
#
# Description
# - Database designed to assign two relays setpoints to a single gauge.
# - Set both relays numbers to same value in case of single relay setup.
# - Off setpoint: pressure in which the gauge generates interlock signal.
# - On setpoint: pressure in which the gauge clears the interlock
# - When pressure reaches OFF level, it must reach ON level so the interlock
# is cleared
# - OFF > ON
#
# Parameters:
# - $(SYSTEM): PV prefix.
# - $(CHANNEL): Channel number with regard to the IOC
# - $(CONTROLLER): Controller name from st.cmd
# - $(PORT): Channel number with regard to the controller
# - $(SP1): 1st setpoint relay
# - $(SP2): 2nd setpoint relay
#

record(ao, "$(SYSTEM):setOffSetpoint$(CHANNEL)")
{
field(DESC, "Sets Cell $(CONTROLLER) Off pressure setpoint")
field(DTYP, "stream")
field(OUT, "@XGS600.proto setRedundantOffSp($(PORT),$(SP1),$(SP2),$(SYSTEM):setOnSetpoint$(CHANNEL)) $(CONTROLLER)")
field(PINI, "NO")
field(PREC, "3")
field(EGU, "mbar")
field(LOLO, "") field(LLSV, "NO_ALARM")
field(HIHI, "") field(HHSV, "NO_ALARM")
}

record(ao, "$(SYSTEM):setOnSetpoint$(CHANNEL)")
{
field(DESC, "Sets Cell $(CONTROLLER) Off pressure setpoint")
field(DTYP, "stream")
field(OUT, "@XGS600.proto setRedundantOnSp($(PORT),$(SP1),$(SP2)) $(CONTROLLER)")
field(PINI, "NO")
field(PREC, "3")
field(EGU, "mbar")
field(LOLO, "") field(LLSV, "NO_ALARM")
field(HIHI, "") field(HHSV, "NO_ALARM")
}

record(ai, "$(SYSTEM):getOffSetpoint$(CHANNEL)")
{
field(DESC, "Reads Cell $(CONTROLLER) Off pressure setpoint")
field(SCAN, "1 second")
field(DTYP, "stream")
field(INP, "@XGS600.proto getOffSetpoint($(SP1)) $(CONTROLLER)")
field(PINI, "YES")
field(PREC, "3")
field(EGU, "mbar")
field(LOLO, "") field(LLSV, "NO_ALARM")
field(HIGH, "") field(HSV, "NO_ALARM")
field(HIHI, "") field(HHSV, "NO_ALARM")
}

record(ai, "$(SYSTEM):getOnSetpoint$(CHANNEL)")
{
field(DESC, "Reads Cell $(CONTROLLER) Off pressure setpoint")
field(SCAN, "1 second")
field(DTYP, "stream")
field(INP, "@XGS600.proto getOnSetpoint($(SP1)) $(CONTROLLER)")
field(PINI, "YES")
field(PREC, "3")
field(EGU, "mbar")
field(LOLO, "") field(LLSV, "NO_ALARM")
field(HIGH, "") field(HSV, "NO_ALARM")
field(HIHI, "") field(HHSV, "NO_ALARM")
}

record(mbbi, "$(SYSTEM):getChannel$(CHANNEL)Relay1State")
{
field(DESC, "Read channel $(CHANNEL) relay 1 state")
field(DTYP, "stream")
field(INP, "@XGS600.proto getSetpointState($(SP1)) $(CONTROLLER)")
field(SCAN, "1 second")
field(PINI, "YES")
field(ZRVL, "0") field(ZRST, "OFF") field(ZRSV, "MAJOR")
field(ONVL, "1") field(ONST, "ON") field(ONSV, "MAJOR")
field(TWVL, "3") field(TWST, "Auto") field(TWSV, "NO_ALARM")
}

record(mbbi, "$(SYSTEM):getChannel$(CHANNEL)Relay2State")
{
field(DESC, "Read channel $(CHANNEL) relay 2 state")
field(DTYP, "stream")
field(INP, "@XGS600.proto getSetpointState($(SP2)) $(CONTROLLER)")
field(SCAN, "1 second")
field(PINI, "YES")
field(ZRVL, "0") field(ZRST, "OFF") field(ZRSV, "MAJOR")
field(ONVL, "1") field(ONST, "ON") field(ONSV, "MAJOR")
field(TWVL, "3") field(TWST, "Auto") field(TWSV, "NO_ALARM")
}
75 changes: 75 additions & 0 deletions ipApp/Db/XGS600.proto
Original file line number Diff line number Diff line change
@@ -0,0 +1,75 @@
#
# XGS600.proto
#
# Stream device protocol file for Agilent XGS600 vacuum gauge controllers
#
# Abdalla Al-Dalleh
# August 13, 2026
#

Terminator = CR;

# $1: Channel Number [1-5]
getPressure
{
out "#0002I\$1";
in "%*1c%E";

@mismatch {
in "%*1c%E";
}
}

# $1: Relay Number [1-8]
getOffSetpoint
{
out "#009\$1";
in "%*1c%E";
}

getOnSetpoint
{
out "#008\$1";
in "%*1c%E";
}

# $1: Channel Number [1-5]
# $2 $3: Relay Number [1-8]
setRedundantOffSp
{
out "#006\$2I\$1%(\$4.VAL)1.3E";
in ">";
out "#007\$2I\$1%1.3E";
in ">";

out "#006\$3I\$1%(\$4.VAL)1.3E";
in ">";
out "#007\$3I\$1%1.3E";
in ">";
}

setRedundantOnSp
{
out "#006\$2I\$1%1.3E";
in ">";
out "#006\$3I\$1%1.3E";
in ">";
}

setSetpointState
{
out "#005E%(\$1)u%d";
}

getSetpointState
{
out "#005F\$1";
in "%*1c%u";
}

debug
{
ExtraInput = Ignore;
out "%s";
in "%50c";
}