Certified Reduced-Lead ECG Reconstruction — a per-feature recoverability certificate with distribution-free calibration.
Targeting the IEEE ICASSP Biomedical Signal Processing track. The certificate and
every theory-validation experiment are CPU-only and fully reproducible on real public
data (PTB-XL); the single generative-fabrication exhibit trains a conditional diffusion
model and needs a GPU (experiments/gpu_diffusion_clean.py). Every theorem in the
paper is cross-checked against Monte-Carlo simulation by the test suite before it is
allowed into the manuscript.
TL;DR — Reconstructing missing ECG leads from a reduced set is an ill-posed inverse problem, and a scalar reconstruction error says nothing about which features a clinician can trust. We output, per waveform feature (P/QRS/ST/T) and per lead, a certificate of what is provably recoverable (with a closed-form conditioning κ that warns, before any reconstruction, which lead configurations can support a feature), what is statistically recoverable (distribution-free calibrated interval), and what is provably unrecoverable (flagged, held-out false-flag ≤ α). The engine is a fact the ECG has and generic inverse-problem theory does not: each waveform segment is an approximately rank-3 cardiac dipole plus a non-dipolar residual. A real conditional diffusion model, scored against a held-out recoverability oracle, shows fabrication is a choice of objective, not an accident of the network: raising the realism knob injects a growing recoverability deficit that the certificate localizes and a scalar RMSE cannot see.
- The problem in plain language
- The idea: three tiers of recoverability
- The theorem (and what is not ours)
- The certified estimator
- Results
- Repository layout
- Installation and full reproduction
- Data access (important under GFW)
- What each test proves
- Relation to prior work
- Honest limitations
- 中文速览
- Citation & license
A 12-lead ECG is expensive to acquire; wearables and monitors often record only one, three, or six leads. Lead reconstruction fills in the rest. Dozens of recent methods — linear, U-Net, transformer, diffusion — report a low mean-squared error (MSE) and declare success.
But a low average error certifies nothing about which features are trustworthy. A cardiologist does not read "average millivolts"; they read the P wave, the QRS complex, the ST segment, the T wave — each on specific leads. If a reconstruction gets the average right but invents an ST elevation on V2, it invents a heart attack.
Generative models are especially dangerous here, and for a principled reason. To look realistic on an ill-posed problem you must add content you cannot know (the perception-distortion tradeoff). On an image that yields a plausible-but-wrong texture; on an ECG it yields a plausible-but-wrong diagnosis.
Write reduced-lead reconstruction as an inverse problem y_S = A_S · L + noise, where
L is the full 12-lead signal and A_S selects the observed leads S. Because the
standard 12 leads obey exact algebra (Einthoven / Goldberger: III = II − I,
aVR = −(I+II)/2, …), the operator A_S is known — we never estimate it.
For each waveform segment s ∈ {P, QRS, ST, T} the instantaneous heart potential is,
to good approximation, a 3-D cardiac dipole M_s·d plus a non-dipolar residual
r_s. We estimate the per-segment dipolar basis M_s ∈ ℝ^{12×3} from a population
(the top-3 left singular vectors of the segment's lead covariance). This splits every
reconstructed feature into three tiers:
| Tier | What it is | Guarantee |
|---|---|---|
| I — recoverable | the dipolar component, when S spans the dipole |
exact up to a closed-form noise gain κ_s(S) |
| II — statistical | population-correlated non-dipolar content | distribution-free calibrated interval |
| III — unrecoverable | non-dipolar content independent of S (local, e.g. a fractionated QRS under one precordial electrode) |
any reconstruction of it is fabrication → flagged with false-flag rate ≤ α |
The number κ_s(S) = ‖M_s M_{s,S}⁺‖₂ is the key object. It says geometry beats lead
count: three coplanar limb leads see only 2 of the 3 dipole directions, and six
frontal-plane limb leads recover the (transverse) dipole far worse than three
well-chosen leads. Measured on PTB-XL QRS:
| configuration | κ |
note |
|---|---|---|
| Lead-I | rank 1 | 2 of 3 dipole directions are Tier III |
{I, II, V2} (spread) |
3.1 | well-conditioned |
{V1, V2, V3} (adjacent) |
4.9 | spans, but worse |
{I, II, III} (coplanar) |
3.9×10⁵ | looks like 3 leads, is really 2 |
| limb-6 | 6.7×10⁴ | 6 leads, all frontal-plane |
Theorem (per-feature dipolar recoverability). With M_{s,S} the observed rows of
M_s: if rank(M_{s,S}) = 3, the population-dipolar projection of every lead is
recovered by L̂_s = M_s M_{s,S}⁺ y_S with error M_s M_{s,S}⁺ n, so the error is
≤ κ_s(S)·‖n‖. If rank < 3, the unobserved dipole directions are unrecoverable at
any SNR. The residual splits into a y_S-predictable (Tier II) and a y_S-independent
(Tier III) part.
This is a positive guarantee — which named feature, on which lead, is
recoverable and with what noise gain — that a generic inverse-problem analysis cannot
produce, because it has no M. That is what makes the work non-derivative rather than
"apply known UQ to ECG."
What we do not claim. The negative side — that Tier III content is unrecoverable
and any estimator returns the prior mean with error ≥ Var(u) — is the standard
non-identifiability limit, established in general form for inverse problems by
[Iagaru & Gottschling et al., arXiv:2605.13146] and [Kim & Fridovich-Keil,
arXiv:2510.10947]. We credit and use it; our contribution is the ECG-physical
instantiation, the positive κ_s(S) certificate, and the clinical metering.
- Tier II — calibrated intervals. Group-conditional (Mondrian) conformalized
quantile regression (CQR): a separate finite-sample conformal correction per group
(segment, lead), so coverage holds conditional on the feature and lead, not just marginally. - Tier III — the flag. The hallucination energy
h_{s,ℓ}is the reconstruction's energy in the certified-unrecoverable subspace. We threshold it at the one-sided(1−α)conformal quantile ofhover faithful reconstructions, so the false-flag rate is≤ α, finite-sample and distribution-free. Conformal Risk Control setsαagainst the clinical loss (a missed STEMI ≫ a false alarm). - Device shift. Under a genuine device shift (Schiller CS-family → AT-family within PTB-XL), Tier I exactness and Tier III soundness are distribution-free by construction; only Tier II coverage level is shift-sensitive, and it is restored by weighted conformal + a small-slice recalibration.
Synthetic validation (results/synthetic_validation.png) — all three claims hold
exactly: Tier I error grows linearly with noise at slope κ; the Bayes-optimal
reconstructor's Tier III error equals the Var(u) lower bound while a hallucinator
doubles it; the flag's false-flag rate is 0.099 ≤ α=0.1 with power → 1.0; Mondrian-CQR
attains 0.90 coverage.
PTB-XL — the certificate localises and explains. Reconstructing from {I, II, V2}
on the held-out fold, a scalar RMSE ranks three reconstructors (the generative model
is worst) but does not say which feature is wrong or why. The certificate does:
| segment | method | RMSE (mV) | h (mV) |
ρ (oracle) |
|---|---|---|---|---|
| QRS | dipolar | 0.196 | 0.000 | +0.00 |
| QRS | OLS (learned linear) | 0.196 | 0.030 | +0.13 |
| QRS | generative | 0.228 | 0.097 | +0.00 |
| ST | dipolar | 0.125 | 0.000 | +0.00 |
| ST | OLS | 0.112 | 0.016 | +0.28 |
| ST | generative | 0.149 | 0.063 | −0.00 |
| T | dipolar | 0.140 | 0.000 | +0.00 |
| T | OLS | 0.105 | 0.028 | +0.44 |
| T | generative | 0.163 | 0.067 | −0.00 |
h is the deployable hallucination energy (no ground truth); ρ is an oracle
diagnostic (needs the true leads) used to validate that flagged energy is fabricated.
The dipolar baseline never fabricates (h=0) but recovers only Tier I; OLS's
non-dipolar energy is correlated with truth (it genuinely recovers Tier II); the
generative model's is not. Honest caveat: h alone need not separate blur from
fabrication when energies match — the dependable, deployable signal is the
configuration-level certificate (κ, tiers) plus the held-out flag.
Fabrication is the objective, not the network — a real diffusion model, measured
non-circularly (results/gpu_diffusion_frontier.png). A subtlety makes the naive test
circular: on the certified-unrecoverable subspace, correlation-with-truth ρ≈0 for any
sampler (even Bayes-optimal), so ρ≈0 there measures the definition, not fabrication. We
defeat this with a held-out recoverability oracle — a supervised ridge predictor whose
out-of-sample correlation ρ_oracle is a lower bound on recoverability: ρ_oracle>0
proves content is recoverable (a linear map attains it). For S={I,II,V1,V3,V5}→{V2,V4,V6}
the QRS non-dipolar content is provably recoverable (ρ_oracle=+0.33). We then train a real
conditional 1-D DDPM (classifier-free guidance + RePaint) on PTB-XL and score it with the
certificate, comparing the diffusion posterior mean (variance-free) to a single deployed
draw to avoid a mean-vs-sample confound. At the honest end (guidance w=1) the posterior
mean matches the oracle (Δρ≈0 — the model is competent); as guidance pushes toward realism,
the mean develops a growing recoverability deficit — monotone in the seed mean and in
3 of 4 independently trained models (all 4 for w≥2), reaching Δρ = +0.15 ± 0.03 at
w=6 (≈4.3σ over seeds). Through the operational regime (w≤4) RMSE stays flat (~0.21
mV, posterior mean) and the model stays faithful on the recoverable dipole subspace
(ρ_recoverable≈0.89) even though the deficit is already present, if marginal (+0.08 ± 0.04, ~1.9σ)
— the fabrication is invisible to a scalar and localized only by the certificate. Only at
extreme guidance (w=6) does the model visibly degrade on every metric (RMSE→0.32), by which
point the certificate had flagged it all along. On the limb-6 control the linear reference is
lower but nonzero; the diffusion matches or exceeds it (Δρ stays negative) — no false alarm. (A
complementary results/neural_perception_distortion.png shows the same phenomenon in a CNN
swept over an explicit distortion–perception weight.)
Safety case (honest). Reconstructing precordial leads from limb-6 (κ=6.7×10⁴,
effective rank 2 → precordial ST is Tier III) is certified unsafe upfront. Over 799
records the danger is bidirectional and RMSE-invisible (~0.075 mV ST error all methods):
the generative model fabricates 263 phantom STEMIs, OLS masks 360 real ones, and
even the dipolar reconstructor fabricates 226 (inside the recoverable subspace, where
h is blind — its flag correctly never fires). The null-space flag catches only a
minority of fabrications (18/263, held-out threshold). The
reliable safety signal is the configuration-level κ warning, not the per-record flag.
Device shift. CS→AT device shift drops Tier II coverage 0.90→0.82; a 300-record recalibration restores it to 0.93 at modest width increase (0.15→0.24 mV). (Naive likelihood-ratio weighting over-widens to trivial 1.00 coverage — not a repair.) Only the Tier II level is shift-sensitive; Tier I exactness and Tier III soundness hold.
Cross-dataset (results/cross_dataset.json). The certificate's only data-estimated
object — the dipolar subspace M_s, hence the conditioning κ_s(S) — is
population-independent. Refit on the geographically distinct Chapman-Shaoxing-Ningbo
database (349 Chinese hospital records, identical processing), the per-segment dipolarity
matches PTB-XL (QRS 0.88/0.91, ST 0.76/0.72, T 0.76/0.74), the dominant QRS dipole directions
align (principal angles ≤21°), and the well-conditioned κ_s(S) agrees (the exhibit's
{I,II,V1,V3,V5}: 2.3 vs 2.5). The lead-mixing operator T (Einthoven/Goldberger) is
dataset-independent by definition; the fit confirms the estimated M_s is too — except
the lowest-variance dipole direction of the less-dipolar ST/T segments, and limb-6, which is
effectively rank-deficient in both cohorts (κ≳10⁴, magnitude numerically unstable). (κ
is not pure algebra — it depends on the estimated M_s; only the lead operator T is.)
src/ecgcert/
physics/dipolar_subspace.py # lead algebra, per-segment M_s, kappa_s(S) certificate
certify/tier_decomposition.py # Tier I/II/III projectors, hallucination energy h
conformal/mondrian_cqr.py # Mondrian-CQR, conformal risk control, weighted conformal
estimators/reconstructors.py # dipolar / Bayes / OLS / generative reconstructors
models.py # fit per-segment (M_s, mu_s, Sigma_r) from a population
clinical.py # ST-deviation measurement + STEMI flip counting
data/ptbxl.py # PTB-XL loader + NeuroKit2 P/QRS/ST/T delineation
estimators/diffusion.py # conditional 1-D DDPM (CFG + RePaint), GPU
experiments/
synthetic_dipole_injection.py # M3: validates all theorems (figures)
ptbxl_reduced_lead.py # M4: 3 configs, baselines, hallucination quantification
ptbxl_stemi_safety.py # M5: fabricated/masked STEMI + abstention (held-out flag)
cross_device.py # M5b: CS->AT device-shift coverage + repair
neural_baselines.py # M6: real CNN distortion->perception sweep (CPU)
gpu_fabrication.py # M7: oracle gate (Band A/B) + diffusion scoring (GPU)
gpu_diffusion_clean.py # M8: de-noised CRAFT exhibit + --ci error bars (GPU)
gpu_diffusion_figure.py # M8: recoverability-deficit frontier figure (+CI bars)
cross_dataset.py # M9: Chapman-Shaoxing-Ningbo M_s/kappa transfer check
scripts/
download_data.py # PTB-XL via HuggingFace mirror (GFW-friendly)
precheck_dipolarity.py # risk-2 gate: per-segment dipolarity by diagnosis
tests/ # 19 theorem-vs-simulation checks
paper/ # ICASSP spconf LaTeX source
# 1. environment (Python 3.11)
uv venv --python 3.11 .venv
uv pip install --python .venv/Scripts/python.exe -e .
# 2. data (PTB-XL, ~1.8 GB, via HuggingFace mirror — see §8)
.venv/Scripts/python.exe scripts/download_data.py --dataset ptbxl
# 3. the risk-2 gate + all experiments
.venv/Scripts/python.exe scripts/precheck_dipolarity.py
.venv/Scripts/python.exe experiments/synthetic_dipole_injection.py
.venv/Scripts/python.exe experiments/ptbxl_reduced_lead.py
.venv/Scripts/python.exe experiments/ptbxl_stemi_safety.py
.venv/Scripts/python.exe experiments/cross_device.py
# 4. tests (all theorems cross-checked by simulation)
.venv/Scripts/python.exe -m pytest -qThe synthetic experiment and calibration run in seconds on a CPU; the PTB-XL experiments are dominated by NeuroKit2 delineation (a few minutes each).
PhysioNet's HTTPS endpoint is unreachable from some networks (direct connections are
dropped; some proxies reset the TLS handshake to physionet.org specifically). The
downloader therefore pulls PTB-XL from the byte-complete HuggingFace mirror
longisland3/ptb-xl via
hf-mirror.com with curl (resumable). Raw data is not committed to git; run the
downloader. If your network reaches PhysioNet directly, use --source physionet.
| test | claim |
|---|---|
test_physics::test_lead_algebra_rank8_and_relations |
the encoded 12-lead algebra is exact (rank 8, Einthoven/Goldberger) |
test_physics::test_dipolar_subspace_matches_dower |
the data-estimated dipolar subspace equals the inverse-Dower column space |
test_physics::test_tier1_exact_recovery_noiseless |
Tier I recovers all 12 leads exactly for a dipole-spanning set |
test_physics::test_kappa_geometry_not_leadcount |
κ / rank distinguish spanning from coplanar/collinear configs |
test_physics::test_tier1_noise_amplification_matches_kappa |
reconstruction error obeys ‖·‖ ≤ κ‖n‖ and attains it |
test_certify::* |
Tier projectors, h = 0 for faithful dipolar, h grows with injection, supported-reconstruction strips fabrication |
test_conformal::* |
marginal + group-conditional coverage, false-flag control, weighted conformal recovers coverage under shift |
test_synthetic_theory::* |
the paper's synthetic figures match their theorems (Tier I κ law, Tier III Var(u) bound, flag, coverage) |
- Reduced-lead reconstruction (U-Net/GAN/diffusion, e.g. arXiv:2502.00559, arXiv:2401.05388): reports aggregate MSE; qualitatively notes non-dipolar content is hard. We turn that into a per-feature certificate with calibration and a flag.
- Hallucination in inverse problems (arXiv:2605.13146, arXiv:2510.10947): general
theory of unrecoverability and distribution-free assessment. We use it and add
the ECG-physical positive certificate
κ_s(S)and clinical metering. - Vectorcardiography / inverse-Dower (Edenbrandt & Pahlm 1988): the classical whole-beat dipole map, which is our Tier I baseline, not our contribution — ours is per-segment, per-configuration, and calibrated.
- Conformal prediction (CQR, conformal risk control, weighted conformal): the calibration layer; the novelty is the physics that tells conformal what to calibrate.
- The dipolar approximation is weakest in some pathologies; we never assume exact rank 3, and the conformal layer makes subspace approximation affect interval width, not validity (decreasing dipolarity correctly widens intervals and raises flags).
- Dipolarity is feature-and-pathology specific; we do not claim a monotone
"disease lowers dipolarity" (it does not — see
results/precheck_dipolarity.json). - PTB-XL has no per-lead ST-elevation millivolt label; the STEMI safety case uses a measured ST-deviation endpoint (J+60 ms, 0.1 mV), stated as such.
- The deployable hallucination energy
hdetects fabrication in the null space; it is blind to error inside the recoverable subspace (a contaminated dipole estimate), which is instead warned by the conditioning κ. The dependable, deployable signal is the configuration-level certificate (κ + tiers) plus the held-out flag; the correlation ρ used in the results table is an oracle diagnostic (needs ground truth) that only validates whathmeasures. - The sampled "generative" baseline is an idealised perceptual endpoint; the honest fabrication evidence is the real trained-CNN perception-distortion sweep. The certificate is reconstructor-agnostic and wraps any method.
- Theorem 1 is honest about scope: recovery is exact only in the purely-dipolar limit; otherwise the observed non-dipolar residual contaminates the estimate, amplified by the same κ (which is why a large κ is a warning, not just a noise bound).
减少导联的 ECG 重建是病态逆问题。现有方法只报一个平均误差,而生成式重建会编造有临床意义的形态(伪造 ST 抬高等)。本项目按 P/QRS/ST/T 逐特征、逐导联给出可恢复性证书:Tier I(偶极分量,可证精确,噪声增益 κ_s(S) 闭式)、Tier II(人群相关非偶极内容,分布无关校准区间)、Tier III(与观测独立的局部内容,任何"重建"都是幻觉→有保证地标记,假标记率 ≤ α)。理论内核是通用逆问题理论没有的 ECG 物理结构:每个波形段≈秩3心脏偶极+非偶极残差。在 PTB-XL 上,RMSE 几乎相同的重建器被证书清晰区分:生成式在"已认证不可恢复子空间"放入大量能量却与真值零相关(自信地幻觉)。所有定理都由测试套件的蒙特卡洛仿真交叉验证(19 个测试)。