Notifications
You must be signed in to change notification settings - Fork 1
All issues
Issue creation is restricted in this repository
Issues
is:issue state:open
is:issue state:open
Search results
deploy.py dies silently (exit 127) after flashing when stdout is piped — post-flash comeback + --power-cycle-cmd recovery never runs
bugSomething isn't workingSomething isn't workingStatus: Open.#121 In OpenwaterHealth/openmotion-sensor-fw;Camera DMA re-arm inside send_data() makes any main-loop stall terminal
bugSomething isn't workingSomething isn't workingStatus: Open.COMM TX busy-wait (TX_TIMEOUT 500 ms) stalls the main loop and kills a camera
bugSomething isn't workingSomething isn't workingStatus: Open.In-app path to write the camera FPGA bitstream to flash at 0x081A0000 (bitstream updates on bootloader units)
featureNew featureNew featureStatus: Open.#112 In OpenwaterHealth/openmotion-sensor-fw;Host USB read stall ends the scan early and triggers ~88 spurious scan start/stop cycles
bugSomething isn't workingSomething isn't workingStatus: Open.#101 In OpenwaterHealth/openmotion-sensor-fw;Drip-scan image receive mode: OW_CAMERA_IMAGE_MODE (0x30) + data-paced 2408 B line forwarding
featureNew featureNew featureStatus: Open.#99 In OpenwaterHealth/openmotion-sensor-fw;DFU boot magic at 0x38000000 collides with HISTO SOF marker — streamed data can spoof a bootloader request across warm reset
bugSomething isn't workingSomething isn't workingStatus: Open.#93 In OpenwaterHealth/openmotion-sensor-fw;Host-upload FPGA SRAM path programs zeros: xi2c_write_long clobbers its own source buffer
bugSomething isn't workingSomething isn't workingStatus: Open.feat: OV2312 timing mode for >40 fps capture (60 Hz mode blocker)
60hz Sampling ModeTabled 60 Hz capture mode feature — see bloodflow-app#327 for the resurrection runbookTabled 60 Hz capture mode feature — see bloodflow-app#327 for the resurrection runbookfeatureNew featureNew featureStatus: Open.feat: 60 Hz frame-rate support — internal FSIN rate select + frame-budget scaling
60hz Sampling ModeTabled 60 Hz capture mode feature — see bloodflow-app#327 for the resurrection runbookTabled 60 Hz capture mode feature — see bloodflow-app#327 for the resurrection runbookfeatureNew featureNew featureStatus: Open.