Ventway Quick Reference

What is this project

Bare metal ventilator controller on STM32F4. No OS, no HAL. Runs in Renode (hardware simulator). PID controls turbine speed to hit pressure targets. Fixed-point math, no FPU.

Peripherals and registers

A peripheral is any hardware block on the chip that isn't the CPU — timers, UART, GPIO. The CPU talks to them through registers. They do their job independently once configured.

Every peripheral is dead until you turn on its clock (RCC — power saving, only turn on what you need). Then you wire pins to the right peripheral — GPIO alternate function config connects the physical pin to the internal peripheral. Then configure each peripheral.

Three init levels:

  1. Clocks — RCC enables for GPIO, TIM2, TIM3, USART2
  2. Pins — PA2 → USART2 TX (AF7), PA6 → TIM3 PWM (AF2). Mappings fixed by datasheet.
  3. Peripherals — all same level hardware-wise, but TIM2 started last (software concern — its ISR uses everything else):
    • USART2: 115200 baud (16MHz / 139). Fast enough for logs within the 10ms tick, slow enough to be universal. Interrupt on RX byte → ring buffer.
    • TIM3 (PWM): prescaler 15 → 1MHz tick, ARR 999 → 1kHz PWM, 1000 duty steps. Inaudible to motor, 0.1% resolution.
    • TIM2 (tick): prescaler 15999 → 1kHz tick, ARR 9 → 10ms period (100Hz). ~25× oversampling of lung dynamics (RC τ=250ms).

Interrupts need two enables: one on the peripheral side, one on the CPU side (NVIC). Miss either and the handler never fires.

TIM2 interrupt (every 10ms)

TIM2 fires → ISR clears the interrupt flag → calls state_machine_tick():

  1. Decrement state timer. If not expired: read pressure from sensor register (0x50000000), run PID, PID outputs duty percentage.
  2. If timer expired: transition to next state (INHALE→HOLD→EXHALE→INHALE). On EXHALE entry specifically — zero the PID integral, zero duty, reset derivative memory. Other transitions don't touch PID.
  3. Back in the ISR: take the duty percentage PID computed, write it to TIM3 (PWM hardware). TIM3 converts that to a real waveform on PA6 that drives the turbine motor.

That's the entire closed loop. Sensor → PID → PWM, 100 times a second.

Main loop

After all peripherals are initialized and TIM2 is started (last, because it kicks off interrupts), main enters an infinite loop:

  1. Check if state changed (flag set by ISR) → if yes, log the transition over UART (state name, pressure, duty).
  2. Drain RX ring buffer — bytes arrive from UART interrupt into a ring buffer, main loop pulls them out and feeds them to the command parser (runtime PID tuning, timing changes).
  3. Flush TX ring buffer to UART hardware — ISR writes log lines into a ring buffer, main loop actually pushes bytes out over serial.
  4. WFI — wait for interrupt. CPU sleeps until next TIM2 tick or UART byte arrives. Saves power, no busy-waiting.

State machine

INHALE (1s, target 20) → HOLD (0.5s, target 20) → EXHALE (1.5s, target 5). Repeat. ~20 breaths/min.

HOLD exists to let flow drop to zero so you see plateau pressure (pure elastic V/C) — clinically important.

On EXHALE entry: zero PID integral, zero duty, reset derivative memory. Other transitions don't touch PID state.

PID

  • P: proportional to error (target - pressure)
  • I: sum of past errors — eliminates steady-state error. Summing not averaging — measures "how long off target"
  • D: derivative on measurement (not error) to avoid setpoint kicks
  • Anti-windup: back-calculation. When output saturates (clamped to 0-100%), the difference feeds back into the integral to pull it down. Prevents integral from growing unbounded during saturation.

Integration testing

Three separate processes, strict boundaries:

  1. ventway — firmware process. Runs on its own (emulated) CPU. Only talks through registers — reads pressure from 0x50000000, writes duty to TIM3. Knows nothing about lungs.
  2. lung_model — patient process. Independent physics simulation. Receives flow, computes pressure (RC circuit). Knows nothing about firmware.
  3. sim — glue process. Connects the other two through the register interface. Reads duty from ventway, feeds it to lung_model as flow, takes pressure back, writes it to the sensor register.

Each process is a separate codebase, separate build, separate unit tests. They can only communicate through the register interface — no function calls, no shared memory, no imports. Process isolation enforces the boundary.

The closed loop: ventway writes duty → sim reads it → lung_model turns it into pressure → sim writes pressure to 0x50000000 → ventway reads it next tick. Same loop as real hardware: ventilator → turbine → air → patient → pressure sensor → ventilator.

Startup: Robot Framework (4th process) orchestrates:

  1. Start sim — sets up the register interface
  2. Start lung_model — connects to sim, ready to receive flow
  3. Start ventway — firmware boots, starts reading registers, control loop goes live

Once all three are up, Robot just watches UART output. Pure black-box: it only sees what the firmware prints over serial, same as a real operator. Tests check state transitions happen, pressure hits targets, duty is zero during exhale.