4D motion analysis and ITV generation

The 4D workflow reproduces, inside the viewer, the pipeline of ITV-based motion studies (e.g. upright-vs-supine STAR target evaluation): per-phase registration, target propagation, centroid motion metrics, target-reference drift and correlation, and ITV volumes.

4D groups (src/fourd.rs)

A 4DCT arrives as one image series per respiratory phase, usually with an average and sometimes MIP/MinIP reconstructions beside them; DICOM stores no node for the acquisition they form, so the viewer reconstructs one:

  • Series are bucketed by (study, modality). A series whose description carries a number directly before a % (e.g. Thorax 4D 30%, CT 0 % Ex) or phase + number (4DCT_phase_000, Phase 3) is a phase; the description minus the number (the template) tells two 4D sets in one study apart - thin- and thick-slice reconstructions become two groups - and is the group's name stem (4DCT - Thorax 4D (10 phases)).
  • Series with a TemporalPositionIdentifier but no percent group by identical description and order by that identifier (t1, t2, …).
  • AVG/average/mean, MIP and MinIP in the description mark the reconstructions; they attach to the bucket's first group.
  • A group needs at least three phases - two series with "50%" in the name are more likely a coincidence than an acquisition.

Detection is a heuristic, so everything can be corrected from the data tree: right-click a series ▸ 4D group to add it to a group or start a new one; a group member to reorder it, change its role or remove it; the group node to rename it, dissolve it or re-run detection. Hand-edited groups are marked custom and survive re-detection (fourd::refresh, run whenever the series list changes); a dissolved auto-detected group leaves a hidden tombstone that only the explicit Re-detect 4D groups clears. Members reference series by SeriesInstanceUID, so renames, removals and copies never corrupt a group.

A group can also be built from a selection when detection has nothing to go on - ten series called CT with no percent, no phase and no temporal index. Ctrl-click ticks series in the tree; right-click a ticked row ▸ 4D group ▸ New 4D group from the N ticked series, or right-click the modality node for every series under it (fourd::group_from). The phases are ordered by percent when every phase has one, else by TemporalPositionIdentifier, else by series number, else in tick order, and a phase with no label of its own is called t<n> by position; reconstructions the descriptions name (AVG, MIP, MinIP) go last. The series taken leave any group they were in, and a group emptied that way is dropped. Add the N ticked series to appends them as phases.

Every engine of the Structure auto tools module can run on every phase of a group in one go (auto-segmentation.md).

In the tree a group renders as a 🎞 node inside its study - phases in temporal order, then the reconstructions; grouped series leave their modality node so each series has one place.

Watching a group move

Any viewport showing a series of a group carries a ▶4D button, and the Playback module carries the same transport with a phase scrubber; both are described in viewer.md. The short version: the first press reads every phase into memory (a group whose phases would exceed the module's budget is refused rather than read), and then the workspace runs through the phases with the structure set, segmentation series and dose of each one, the tree selection walking down the group as it goes. Stepping between the phases of the group on display keeps the crosshair, the zoom, the pan and the registration where they are, which is what makes two phases comparable by eye. The 3D window plays the same run and can mesh every phase up front so the surfaces keep up.

Playing is for looking; the pipeline below is for measuring.

The pipeline (src/app/motion_win.rs)

Tools ▸ 📈 Structure motion (the workspace is chosen on the window's Workspace A / B row), or right-click a 4D group ▸ Motion / ITV analysis…. One run:

  1. Reference phase - chosen in the dialog (default: the 0 % phase). Targets are defined on it: contours rasterized onto its lattice, segmentations from another lattice resampled onto it.

    The Targets list groups structures by name across the phases. The first column, On every phase, holds names that exist on each phase of the group (a target propagated onto every phase's structure set, say); one tick there is one target, not ten. The second column, On some phases, lists the rest with the series they belong to; such a target is read from the reference phase alone.

  2. Per-phase registration - the reference volume is registered to every other phase with the elastix engine. Settings (levels, iterations, samples, grid spacing, sampling threshold) come from the Registration panel, adjustable in the dialog. The models:

    • rigid is, by default, a local rigid fit per structure: the region registered is the structure dilated by the local margin (15 mm by default), so the fit follows that structure and not the whole image. Set the margin to 0 for one global rigid fit of the whole volume. A global fit of a breathing thorax is dominated by the spine, ribs and couch, which do not move, and reports a motion of zero for everything; that is what the local fit is for.
    • deformable is a B-spline refinement started from the global rigid result.
    • as contoured uses no registration at all: the structure is read from each phase's own contours, so the track is exactly what was drawn (or propagated) on each phase. It is offered when a ticked target exists on every phase and is then the reference the others are judged against.
  3. Propagation - every target (and the reference structure, when chosen) is carried through each transform onto each phase's lattice, once per model; the transform maps reference → phase, so landing on the phase samples through the inverse, as in the propagation tool. As contoured rasterizes the phase's own contour instead.

  4. Measurement (src/motion.rs) - per phase and model: centroid (mm, patient LPS), volume; from those: displacement from the reference phase, 3D magnitude, peak-to-peak amplitude (largest pairwise centroid distance), target-reference drift |TV − ref| and its peak-to-peak, and Pearson correlation of target vs. reference displacement along RL / AP / SI with two-tailed p-values (t-test, n−2 dof).

  5. ITV - one per target and model, the union of the propagated masks over the selected phases, resampled onto the reference lattice, plus an optional uniform margin. The Phases row next to Build ITV (All / None, and the Select phases list under it) chooses which phases take part in the run; the reference phase always does. ITVs land as a segmentation series 4D ITV - <group> referencing the reference phase series (display that phase to see and edit them; they export like any segmentation - SEG or RTSTRUCT).

  6. Registration QA - per phase and model: the Dice of the two images' tissue after the registration against what it was before it, the engine's metric line, the 95th-percentile displacement, and the folding rate (fraction of sampled points with a non-positive Jacobian). The Dice leads every row and is coloured by the usual bands (0.80 and above green, 0.60 to 0.80 amber, below red), so a phase that did not land on the reference is visible without reading the metric line. Where a phase carries its own contour of a target, the propagated structure is also scored against that contour and the score is listed under the phase (GTV vs contoured): the honest measure of whether the model followed the anatomy, not just the image. Both go into the CSV's Registration quality section (Image Dice after / before, Dice <structure> vs contoured) and into the MCP report as image_dice and structure_dice.

Keep per-phase segmentations additionally stores every propagated mask as a segmentation series on its phase (4D <phase> - <group>).

Cancel stops the run at the next phase boundary; a finished run is never applied to a workspace that was replaced while it ran.

Recipes - several studies, one workflow

Starting a run remembers the dialog as a recipe: target and reference-structure names, models, ITV options and registration settings. Apply last recipe re-ticks the same structures by name in whatever workspace the dialog is open on - load the next patient (or the paired upright/supine study into workspace B), open the tool, apply, run. Recipes are name-based on purpose: indices and UIDs do not travel between patients.

Results (src/app/motion_results.rs)

The results window opens when a run finishes (later: Tools ▸ 📈 Motion results). Per run: the displacement-magnitude-vs-phase chart (one line per target and model, reference structure dashed red), peak-to-peak amplitude and drift bars (one per target and model), the per-phase table (|d| and volume per track), correlation lines (r, p, significance stars, synchrony wording), registration-QA lines and ITV volumes. A target that was ticked once is one line and one amplitude; ticking the same name on every phase separately would have made ten targets of it, which is what the grouped Targets list prevents.

The window lays itself out to its own size: the run pickers wrap onto a second line when the window is narrow, the export buttons stay at the bottom, and only the report scrolls, vertically. The chart takes the full width and scales its height with it (never more than about 60 % of the visible height), and its margins are measured from the labels that go in them - the widest value label on the left, half the last phase label on the right - so the axis labels and the last phase are always on screen. A per-phase table wider than the window scrolls sideways on its own.

The value axis uses a step of 1, 2 or 5 × 10ⁿ - the one whose interval count is nearest to what the chart's height has room for (about one tick per 34 points) - rounded out to whole steps, with as many decimals as the step needs (a 0.5 mm step reads 0.0 0.5 1.0 1.5 2.0), so no two ticks carry the same label. The axis spans at least 1 mm, so a track that hardly moves stays flat instead of being blown up to fill the chart. Hovering the chart marks the phase under the pointer and lists every line's value there.

Compare with puts a second run beside the first - workspace A vs. B, upright vs. supine - matching ITVs and tracks by target name and model: ITV volumes with percentage change, peak-to-peak amplitudes, side by side.

Export CSV writes the selected run as a table a spreadsheet opens as it is: the phases are the columns and every method has its own row, the methods of one quantity on consecutive rows, so rigid, deformable and as-contoured read against each other phase by phase. The file has a short header (run, workspace, reference phase, reference structure, the axis convention), then three sections separated by a blank line, each with its own header row:

Section Label columns Rows Values
Per-phase values Structure, Quantity, Unit, Method per structure (targets, then the reference structure as <name> (reference)) and quantity, one per method one column per phase
Summary Structure, Method one per structure and method peak-to-peak, largest |d|, target-reference drift peak-to-peak, correlation r and p per axis, ITV volume, margin and name
Registration quality Quantity, Unit, Method, Fit per quantity, one per method and fit (whole image, or <structure> neighbourhood for a local rigid fit) one column per phase; the reference phase is not registered and stays empty

The per-phase quantities are the centroid position (Centroid RL (x), AP (y), SI (z), mm, patient LPS), the displacement from the reference phase per axis and as Displacement |d|, the volume (cm3), the grey-level minimum, mean and maximum inside the structure (where the phase images were at hand), and for a target, where the same method also carried the reference structure, Offset to <reference> RL / AP / SI (target centroid minus reference-structure centroid). The registration-quality quantities are the image Dice after and before, the metric at start and at end (its unit column names the metric, MSD or MI), iterations, time, the 95th-percentile displacement, the folding rate and, where a phase carries its own contour, Dice <structure> vs contoured.

Everything in the file is readable in any program: the typographic symbols the window uses are written in plain ASCII (· as -, ▶ as ->, cm³ as cm3), and a file that still holds a letter outside ASCII (a structure named in another alphabet, say) starts with a UTF-8 byte-order mark, which is what makes Excel read it as UTF-8 rather than in the system's code page. Names with a comma are quoted. Export comparison CSV writes both runs one after the other and then a Comparison section: one row per structure and method, with the two peak-to-peak amplitudes and their difference, and the two ITV volumes and their change in percent. The MCP motion_report tool returns the same CSV text.

Transfer by relationship (src/app/transfer_win.rs)

Tools ▸ ◎ Transfer by relationship… places a structure of one workspace into the other at the same offset from a reference structure's centroid - the STAR workflow's target-heart relationship: a target is projected into a workspace registration cannot reach (another patient, another posture) via anatomy both workspaces can segment. The target keeps its shape; the tool reports the offset (RL / AP / SI) it applied. Reference structures whose name contains "heart" are pre-picked.

The window carries the same Transform matrix as the registration module (registration.md), and it shows the relationship's own answer: pick the two reference structures and the grid fills with the shift between their centroids, which is exactly what ▶ Transfer would apply. Ticking Use this matrix takes that over - nudge a number, or put a rotation into it - and places the structure through the numbers instead of the centroid-to-centroid offset, so a rotation can be put into a placement the relationship alone only shifts, and a transform worked out elsewhere can be applied to anatomy the two workspaces do not share. The two reference structures go quiet while it is on, because nothing then asks them anything: only the target is needed, and the result is named <target> (matrix) rather than after a reference it was not placed against.

Structure comparison (src/app/compare_win.rs)

Tools ▸ ◑ Structure comparison computes, for any two structures (either workspace, contours or segmentations, different lattices): volumes, centroid offset (vector and magnitude), Dice, 95th-percentile symmetric Hausdorff distance and the surface distance as mean, SD and maximum. The second mask is resampled onto the first's lattice through patient coordinates; across two frames of reference the window notes the comparison assumes corresponding coordinates. Save CSV writes the lot as a two-column table.

The rigid offset is the one genuinely new computation and has its own tick, because it costs a second on a large structure. The two surfaces carry no point correspondence, so one is invented and then improved: every point of A is paired with the nearest point of B, the rigid body that best explains those pairs is fitted (orthogonal Procrustes, the same registration::analysis::fit_rigid the registration analytics use), A is moved, and the pairing is done again. The fit is always taken from A's original points, so the iteration refines one global transform instead of accumulating a chain of small ones. What comes out is a translation, three Euler angles, and the surface distance left over afterwards next to what it was before - which is how one sees whether a rigid body explains the difference at all.

Two points of interest are compared as points: the window reports their separation instead of a Dice score, which is the target registration error when the two are meant to be the same landmark.

Everything the window measures is shown as a two-column table - quantity and value, the same rows Save CSV writes - with only the warnings and the "why there is nothing to compare" messages left as sentences.

Structure details (src/app/stats_win.rs)

Tools ▸ 📋 Structure details is the other half: one row per structure of one workspace rather than two structures against each other. The columns are the structure's Format (contours, voxels or a point) and its volume twice over - Surface-based, inside a closed triangle surface reconstructed from the contours, and Voxels-based, the voxels the contours fill - which disagree by a few per cent on a coarse series, more on a small structure, and neither of which is wrong; then the grey levels inside the structure (which is how a mis-drawn organ gives itself away), what the geometry costs in slices and points, and whether a derived structure still matches its recipe. Points of interest show their coordinates instead of a volume. How each column is calculated, formula by formula, is in volumes.md. The CSV carries them as surface_based_cm3 and voxels_based_cm3.

The Dice column measures every row against a reference chosen above the table: the structure or segment of the same name in another workspace - one entry per open workspace, so with three of them the table can be measured against whichever is the comparison; it is the default as soon as a second one is loaded, and it is what a propagation, a phase or a second observer is checked with - or one structure or segment of any workspace for all rows (an auto-segmentation against the manual one). Contours are rasterized onto the row's own lattice, a reference on another lattice is resampled onto it, and a row with no counterpart shows -; the tooltip names what it was measured against, and the CSV carries both. The table is computed on demand rather than every frame, and says so when the geometry has changed under it. Save CSV writes it out.

Numerics worth knowing

  • Centroids are exact under the affine lattice→patient map (mean index, then map); peak-to-peak is the largest pairwise distance, independent of the reference-phase choice.
  • The p-values come from the regularized incomplete beta function (continued-fraction evaluation) - the exact t-distribution tail, not a normal approximation; with 10 phases n is small enough for that to matter.
  • HD95/MSD use the exact anisotropic Euclidean distance transform (morphology::dist2_to_foreground) evaluated on surface voxels of each mask against the other, both directions pooled for the percentile.
  • ITV volumes inherit every caveat of nearest-neighbour resampling between phase lattices; centroid metrics are the primary motion descriptors, as in the underlying study design.