Lab Equipment Controller — Specification

What the software is meant to do and the rules it must keep to. This is the durable document: it should stay true release to release.

  • README.md — what it is and what it looks like; each program's own README (Desktop, Core, Cli, Web, Tests) has the build, run and test instructions for it.
  • ARCHITECTURE.md — how the pieces that satisfy this document fit together.
  • UI-SPEC.md — what is on screen and where, in both builds: the control inventory, the metrics, and the rule that the web is a port of the desktop rather than a redesign of it.

1. Purpose and scope

A Windows desktop application for finding and driving bench instruments — oscilloscopes, function generators, multimeters — over Ethernet or RS-232, using SCPI. It replaces per-vendor utilities and a VISA runtime with one small app that speaks the wire protocols directly.

In scope: LAN discovery, connecting to several instruments at once, sending SCPI and reading replies, instrument-aware shortcuts, small scripts, screen and trace capture, export of what it finds.

Out of scope, deliberately:

  • USB/USBTMC and GPIB. Not for the same reason each, and §17 records both. Serial was on this list until it was built (§13); it is here as the one that turned out to be worth revisiting, which is what §17 said about it while it was still an argument.
  • Any dependency on NI-VISA, IVI, or a vendor SDK. The transports are implemented here.
  • Instrument emulation, calibration workflows, or measurement automation beyond scripts.
  • Inventing SCPI. Every command shipped in a catalog or a quick-command button is transcribed from the vendor's programming guide (§10).

2. Platform

Language / UI C# on WinForms (not WPF/XAML)
Target net10.0-windows; the UI-free Core library is plain net10.0
Runtime 64-bit Windows 10/11
Distribution Self-contained single-file .exe (~48 MB), or framework-dependent
Dependencies No VISA, IVI or vendor SDK — the transports are implemented here. In Core, two packages: PdfPig (reading text out of datasheets, §11b) and System.IO.Ports (the serial transport, §13). In the desktop app, WebView2 (the guide viewer, §11a)

3. Concepts

Term Meaning
Device An address found by a scan (ScpiDevice): IP, port, transport, *IDN? reply
Transport How bytes reach the instrument: raw TCP socket, VXI-11, or a serial port (IInstrumentClient)
Identity The instrument's *IDN? string, Manufacturer,Model,Serial,Firmware
Family Broad class inferred from the identity: oscilloscope, Siglent generator, SCPI generator, multimeter, generic
Profile What a family implies: quick-command buttons, capture support (InstrumentProfile)
Session One open connection and its state: client, identity, profile, history, timeout (InstrumentSession)
Console The UI for one session, in a tab or its own window (InstrumentConsole)

4. Discovery

Interfaces. Offer every operational, non-loopback IPv4 interface with a netmask. Order them so the first is the best guess: interfaces with a default gateway first, then by ascending subnet size. A Hyper-V "Default Switch" (/20, no instruments) must not be selected ahead of a real 254-host LAN.

Hosts. Enumerate the subnet that owns the selected address, excluding the network and broadcast addresses, capped at 4096 hosts. Report when the cap truncated the range.

IP range. An optional field narrows the sweep. Empty means the whole subnet, so the default behaviour is what it always was. It accepts 192.168.1.20-60, a bare 20-60 against the selected interface's subnet, a single address, a /28-style block, and any comma-separated mixture of those. A range that cannot be read stops the scan and says why — falling back to the whole subnet because of a typo is the opposite of what was asked. Overlapping ranges probe each host once (§13: these instruments wedge on a second connection). The range in force is named in the status line, so a narrowed scan that finds nothing cannot be mistaken for an empty network.

It exists for a reason smaller than speed: a bench sits at a handful of known addresses, and sweeping all 254 knocks on every printer, camera and PLC that shares the lab network.

Ports. A user-editable list. 111 means VXI-11 (the RPC portmapper); every other port is probed as a raw socket. Defaults offered: 5025 (LXI/VISA convention), 5555 (Rigol), 111.

Probing rules — these are correctness requirements, not tuning:

  • Ports on one host are probed sequentially, never concurrently. Instruments such as the Rigol accept a single connection and wedge if two are opened.
  • A raw-socket probe is a TCP connect only. It must not send *IDN?: on a Rigol an unread reply permanently poisons the output queue (§13).
  • The identity comes from the VXI-11 probe, which is framed and safe to read.
  • Up to 128 hosts are probed at a time.
  • Probe timeouts are fixed and short (connect ~300 ms, identify ~3 s) and are independent of the user's Timeout field, which governs instrument communication only. A large communication timeout must never slow discovery.

Results. One row per address, not per endpoint: all probes for an address are merged, preferring VXI-11 when the instrument offers both, and grafting on an identity read by whichever probe obtained one.

Reporting. Devices are streamed to the UI as they answer, and progress is reported at most ~200 times per scan — a per-host report would swamp the UI thread on a /20 and delay the Stop button behind the backlog. A scan is cancellable, and results found before the stop are kept.

Serial ports are swept differently, and the difference is the physics rather than a preference. Everything above rests on asking every address the same harmless question and believing the answers. A serial port cannot be asked anything until its baud rate, framing and flow control already match, and what is on the other end may be a printer, a modem or a 3D printer rather than an instrument. So the serial half of the scan card (SerialScanner) splits the two halves of "discovery" that a subnet sweep gets to run together:

  • Listing opens nothing. SerialScanner.List(), over SerialPorts.Names() — the CLI's lec ports, the web's GET /api/ports and /api/serial/list, the desktop's port dropdown and its results list. This is what the card shows the moment it switches to Serial, and it is already enough to pick a port and connect. Nothing on anyone's bench is touched by looking at this list.
  • Probing opens ports, and only on a button press. SerialScanner.ScanAsync() opens each port the user chose, at each baud rate the user named, in order, stopping at the first that answers *IDN?. It does not sweep parity, data bits or flow control: that is a combinatorial search against hardware that cannot say "wrong number", and it is the thing §17 refuses. What it does sweep is disclosed on the button, because opening a port asserts DTR and RTS and sends text to whatever is there.

Two further differences from a subnet sweep follow from a port being a thing rather than an address. Every port stays listed, answered or not — an IP that does not answer is simply not a row, but a serial port that does not answer is still a port on this machine and still connectable at settings the scan did not try. And a reply is checked before it is believed: TCP either delivers the sender's bytes or delivers nothing, while a UART at the wrong baud rate delivers different bytes, so a reply is accepted as an identity only if it is printable and comma-separated. One that arrives and fails that test is reported as noise at those settings, which is the fact that sends someone to the baud rate rather than to the cable (§5, §13, §17).

5. Addressing and connection

The Address box accepts, in this order of precedence:

  1. A VISA resource stringTCPIP0::192.168.1.19::inst0::INSTR (VXI-11), TCPIP0::192.168.1.19::5025::SOCKET (raw), or ASRL3::INSTR / ASRL::COM3::INSTR (serial). Parsed natively so a user can paste what NI-MAX or Connection Expert reports. TCPIP and ASRL are the interfaces there are transports for; a GPIB or USB resource string is refused by name, with the reason — it used to fall through to the host branch and be accepted as a hostname, which got the user as far as a DNS failure for an instrument sitting on their desk.
  2. vxi://host — VXI-11 without writing a resource string, and vxi://host/gpib0,9 for an instrument behind a gateway. tcp://host says raw socket the same way.
  3. serial://COM3, or serial:///dev/ttyUSB0 — a serial port, with the line settings as a query when they are not the usual ones. In the GUI the scan card's heading switches to Serial Scan and the address box becomes a list of the machine's ports, so the scheme need not be typed (UI-SPEC.md §3.2); everywhere else it is written out: serial://COM3?baud=115200&parity=even&databits=8&stopbits=1&flow=rtscts&term=crlf. A bare number is the baud rate, so serial://COM3?115200 means what it looks like. Anything unrecognised is refused rather than defaulted: connecting at 9600 to an instrument set to 115200 does not fail, it answers with rubbish. Defaults are 9600-8-N-1, no flow control, LF. The scheme is not optional — COM3 on its own is a legal hostname and cannot be told from a machine on the network.
  4. host:port — the explicit port wins.
  5. A bare host — the port and transport of a matching discovered row are reused, so a previously found address still connects the right way; with no matching row it takes the first port the Port(s) list offers.
  6. Nothing typed — the selected row in the device list.

An explicit port of 111 means VXI-11. A form that names its transport is taken at its word — a discovered row never overrides it, because saying it is the reason to type it. For serial there is nothing to override it with: the device list is what a subnet sweep found, and a subnet sweep does not find COM3 (§17).

One parser, for all three front ends. The desktop, the CLI and the web server read a typed address through the same InstrumentAddress in Core. They did not: the CLI and the server each carried a copy of it and the desktop carried neither, so its box refused vxi:// though UI-SPEC.md §3.3 describes both builds' boxes as taking one.

On connect: open the transport, query *IDN? (falling back to what the scan learned if the instrument won't answer now), derive the profile, and open a console.

One session per address. Connecting to an address that already has a session must focus that console rather than dial again — see §13, the Rigol wedges on a second TCP session. This is a hard rule, not a convenience.

A timeout is not a cancellation. Both cut an attempt short through the same cancellation token, so every deadline the clients impose goes through Deadline, which raises TimeoutException when its own clock is what fired — naming the host, the port and the wait. That covers connecting, reading a binary block over a raw socket, and any VXI-11 RPC call (which also names the procedure: a stalled create_link and a stalled device_read want different things from the user). Only Cancel or Stop produces "cancelled" — a word that tells the user they stopped something. When both land in the same instant, cancellation wins.

A reply that never ended is a failure, not a short answer. A text line that runs out of time throws rather than returning what arrived — this once returned the fragment, which is harmless for an empty reply and dangerous for a partial one: +8.39 in place of +8.39319298E-04 is a plausible voltage, is recorded and plotted as one, and leaves the rest of the line in the buffer to corrupt the read after it. A binary block throws for the same reason, a screenshot cut short being one that comes back looking complete.

Disconnect returns the instrument to local (front-panel) control before dropping the link. This applies on explicit disconnect and on application exit, where each close runs on a worker thread with a short bounded wait so a powered-off instrument cannot hang the exit. A raw socket does this by closing; a serial port has no connection whose closing means anything, so it asks in band with SYSTem:LOCal (§13).

6. Sessions, tabs and detached windows

  • Each connection has its own session and its own console: log, command history, timeout, script editor, and command-reference window. Nothing is shared between instruments.
  • Consoles live in a tab strip; the scan panel and the discovered-instruments list stay shared above them. The tab is captioned Model (address).
  • The main window carries a menu bar. Tools does things to the bench — multi-instrument scripts, the AI connection. Help explains things: the command library, the script language reference, About. Both references are reachable with nothing connected, because both are wanted before there is anything to connect to.
  • The About box reports the version, the runtime and the catalog totals, all read at runtime — the command count is counted from the embedded catalogs rather than written down, so it cannot drift from what ships.
  • Export Results sits immediately after Scan: it exports what the scan found.
  • Any console can be detached into its own window and later re-attached — by its own button or the tab's context menu. Closing a detached window re-attaches it; it never silently disconnects.
  • Detaching reparents the same control. The session, log and history must survive it untouched.
  • Closing a console disconnects its instrument (§5) and removes only that tab.
  • Every tab carries an . It is the discoverable way to reach the close that the tab's context menu already offered, and it acts on mouse up: a press that starts on the glyph and slides off must not disconnect an instrument.
  • Multi-Instrument Scripts sits at the right-hand end of the tab strip. That strip is the list of instruments a script addresses, so the button belongs beside it — and it stays put when there are no tabs, which is exactly when someone might open it to see what a script would need.
  • A console's controls are grouped by what they act on, top to bottom: the identity line and Detach (this console's window), then the quick commands and the tools that ask the instrument for something, then Clear Log / Save Log sitting directly on top of the log they act on, then the command box. Disconnect ends the session, so it sits at the far right of that log strip — away from Detach, which it has nothing to do with, and away from anything clicked routinely.
  • The Timeout field applies to connections that are already open as well as the next one.
  • When no console is present the area explains what to do instead of showing a blank panel.

7. Command console

  • Type SCPI, press Enter (or Send). A command containing ? is treated as a query and its reply is read; anything else is written and acknowledged with (sent).
  • Output is colour-coded: sent command, reply, informational note, error. A failed command reports the error in the log; it never throws away the console.
  • History: Up/Down recall earlier commands for this instrument. The position sits one past the newest entry, so the first Up recalls the last command and stepping back down past the newest clears the input. Blank commands are not recorded.
  • The console is locked out while a script drives the same link — two request/response streams on one connection would interleave and corrupt each other.

8. Instrument families and quick commands

Classification from *IDN?, in this order (the order is load-bearing):

The model is normalised first: a leading MODEL is dropped and spaces and hyphens are removed, because a Keithley answers KEITHLEY INSTRUMENTS,MODEL 2450,… and a Keysight scope MSO-X 3054T — neither would match a prefix test as printed.

Order Family Matched on
1 Tektronix scope maker Tektronix and TDS, DPO, MSO, MDO, DSA, LPD
2 Keysight scope maker Keysight/Agilent/HP and any oscilloscope prefix
3 R&S scope maker Rohde&Schwarz/HAMEG and RTB, RTM, RTA, RTO, RTE, RTP, RTH, HMO
4 Siglent scope maker Siglent and SDS and an X in the model — else Generic
5 GW Instek scope maker GW Instek/Good Will and GDS
6 Fluke multimeter maker Fluke and 884x, 8808, 45
7 Keithley SMU maker Keithley and 24, 26, 6221, 6430, 6514, 6517
8 Keithley multimeter maker Keithley and DMM, 20, 21, 27
9 R&S spectrum analyzers maker Rohde&Schwarz and FPC, FSL, FSW, FSU, FSP (not FSPN), FSQ, FSIQ, FSV/FSVA — a separate catalog each, from that line's own manual
10 (Generic) maker Rohde&Schwarz and any other analyzer prefix (FSE, FSPN) — no catalog
11 Spectrum analyzer DSA, RSA, SSA, SVA, FPC, FSL, FSV, FSW, N90, MS2
12 Multimeter SDM, DM, 344, 34401, 2000, 2110
13 B&K electronic load maker B&K Precision and any electronic-load prefix
14 Chroma electronic load maker Chroma and 632
15 Chroma modular load maker Chroma and 636
16 (Generic) maker Chroma and any other load prefix (the 638 AC loads) — no catalog
17 Electronic load DL, SDL, EL3, N33, IT8, 63, 86
18 Chroma power supply maker Chroma and 62
19 Keysight power supply maker Keysight/Agilent/HP and any power-supply prefix
20 R&S power supply maker Rohde&Schwarz/HAMEG and any power-supply prefix
21 B&K power supply maker B&K Precision and 920
22 B&K triple-output supply maker B&K Precision and 913
23 Power supply DP (not DPO), SPD, E36, PWS, HMP, HMC8, NGE, NGL, NGM, NGP, 22xx
24 Generator SDG, DG, AFG, 33 — split by maker: Siglent → its own dialect, else standard SCPI
25 Oscilloscope DS, MSO, SDS, TDS, MDO, DPO, DSO
26 Generic anything else — IEEE 488.2 common commands only

Rows 10 and 16 route to Generic on purpose. An R&S FSE and a Chroma 63800 would otherwise fall through to the generic analyzer and load catalogs, which are the Siglent SSA3000X and SDL1000X sets. Those partly work, which is the problem: the buttons appear, some succeed, and the failures read as the instrument misbehaving. A vendor-specific test that matches and then declines is the only way to stop a later generic test claiming it.

Row 16 is now permanent as far as anyone can tell. The Chroma 63800 guide has been found and is a scan with no text layer, so there is nothing to extract from it — see §10.

The order is load-bearing and every collision below is real:

  • Vendor-specific scopes first, matched on the maker as well as the model, because the prefixes are shared but the dialects are not. A Tek wants CH1:SCAle and ACQuire:STATE RUN; a Rigol wants :CHANnel1:SCALe and :RUN; an R&S wants CHANnel1:SCALe and a bare RUN, with no leading colon anywhere; a Siglent wants :CHANnel1:SWITch and :TRIGger:RUN. MSO belongs to Rigol, Tektronix and Keysight alike, SDS to Siglent, and DSA is a Tektronix scope but a Rigol spectrum analyzer. Tek's RSA is deliberately not in that list: unlike the DSA it really is a spectrum analyzer.
  • Siglent's scope before its generator is not the concern — SDS and SDG do not collide — but the SDS check must precede the generic oscilloscope test, or a Siglent scope inherits the Rigol catalog through the shared SDS prefix, as it used to.
  • Keithley before the generic multimeter test — a DMM6500 would otherwise be claimed by the DM prefix and given the Siglent SDM's much smaller catalog.
  • Vendor supplies before the generic one — the four address channels four ways: a Rigol names one per command (:APPLy CH1,5,1), a Keysight uses a channel list (VOLTage 5,(@1)), an R&S selects one first (INSTrument:NSELect 1) and then sends unqualified commands to it, and a Chroma is single-output so names none at all.
  • B&K loads before the generic load test86 is in that generic prefix list precisely because of the B&K 8600, so its own catalog has to claim it first. Chroma's loads are 63xxx: the 632xx models take the transcribed 63200A catalog, and the rest decline to Generic rather than inherit the Siglent SDL1000X set.
  • The two B&K supply lines are separate families — same maker, two guides, two command sets. The 9200B is single-output; the 9130B is a triple-output whose commands nearly all act on a selected channel (INSTrument CH2), which the 9200B has no equivalent of. Handing either one the other's catalog would produce buttons that partly work.
  • Spectrum analyzers before scopes — a Rigol DSA815 would otherwise be taken for a scope by the DS prefix.
  • Multimeters before generators — a Siglent SDM multimeter takes standard SCPI and must never inherit the C1:BSWV dialect from its own maker's SDG.
  • DPO excluded from the power-supply test — a Tektronix DPO4104 is a scope, and the Rigol DP800's DP prefix would otherwise claim it.

Tests pin each of these.

A Keithley 2450 or DMM6500 can be set to speak either SCPI or Keithley's own TSP, and answers none of its catalog until *LANG SCPI has been sent and the instrument rebooted. The catalogs carry *LANG, and the script examples lead with *LANG?.

Each family supplies a button set that actually works on it (a scope's :MEASure:VPP? is meaningless to a Siglent generator), plus whether screen capture and waveform capture are offered. An unrecognised instrument still connects and is fully usable from the command line — it just gets the safe generic button set.

9. Scripting

A deliberately small line-oriented language, one instruction per line:

Form Meaning
*IDN? etc. A SCPI line. Contains ? → query and read the reply; otherwise write
# text, // text Comment
DELAY <ms>, WAIT <ms> Pause
PRINT <text>, ECHO <text>, LOG <text> Write a message to the output
REPEAT <n>END Repeat a block; may be nested. ENDREPEAT is accepted for END

Rules:

  • REPEAT with a count of zero or less skips the whole block.
  • END without a matching REPEAT is an error and stops the script.
  • The script stops on the first command error, reporting the offending line number.
  • Cancellation (the Stop button) is honoured between every line and during a wait.
  • A hard instruction ceiling (1,000,000) guards against a pathological script; the Stop button is the normal way out.
  • Scripts are plain text, saved and loaded as .scpi/.txt. An editor belongs to one instrument, named in its title bar.
  • The bundled examples match the instrument the editor was opened for — a multimeter is never offered C1:BSWV, a scope never offered MEASure:VOLTage:DC?. Same sourcing rule as §10: every command is transcribed from that family's guide, so where a guide documents nothing (no error query for the Siglent SDG or SDM; the Rigol spells it :SYSTem:ERRor:NEXT?) the example leaves it out. Any example that enables a generator output must warn on the line above — it can damage whatever is wired up.

9a. Multi-instrument scripts — one script, several instruments

A multi-instrument script drives more than one instrument at a time. The window is Multi-Instrument Scripts; the language, the runner and the .seq files it saves still say sequence for one of them, which is what the code is named after.

It exists because the measurements that need it are interleaved rather than sectioned: a swept filter response sets a frequency on the generator, waits, reads the meter, and repeats — two instruments alternating inside one loop, which a script belonging to one instrument cannot express at all.

It is the §9 language plus five forms:

Form Meaning
DEVICE <alias> : <model> Bind a name to a connected instrument
<alias>: <command> Send this line to that instrument
WITH <alias>END Set the target for a block
FOR <v> = <a> TO <b> STEP <n>END Sweep a value; POINTS <n> [LOG] instead of STEP
<alias>: <query>? -> <name> Capture the reply, used later as $name
RECORD <a>, <b>, … Append a row of results
COLUMNS <a>, <b>, … Name the result columns

Rules:

  • A line with no target is refused when more than one instrument is declared. Guessing would send a generator's command to a meter, which is the failure §10 exists to prevent. With exactly one declared instrument a prefix is unnecessary.
  • DEVICE binds by the model from *IDN?, matched exactly and then by prefix — an SDS2354X answers SDS2354X Plus, and the short name should find it. An ambiguous prefix resolves to nothing rather than to whichever instrument connected first. An address is accepted too, for an instrument that will not identify itself.
  • Every DEVICE is resolved before the first command is sent. A sweep that dies three lines in has already changed the instrument's state.
  • A sweep accepts engineering suffixes (1k, 2.5M). POINTS n LOG spaces points per decade, which is how a filter response is read — a linear sweep from 100 Hz to 100 kHz puts almost every point above 10 kHz and skims the corner.
  • An unknown $name is left as written rather than blanked: FRQ,$typo becoming FRQ, is a command an instrument may well accept, carrying a value nobody chose.
  • RECORD's columns are the commas written in the script, not the commas in the data. The line is split first and each field substituted afterwards, so a captured *IDN? — four commas in every one of them — stays a single column instead of silently becoming five. A field left empty by a reading that never arrived is kept as a gap, because dropping it slides every later value one column left and turns missing data into wrong data; a field left empty by a trailing comma is punctuation and is dropped.
  • -> on a line without ? is an error — there is no reply to capture.
  • Lines run in order. Two instruments never run concurrently: the measurement is inherently ordered, and a connection carries one conversation at a time (§6).
  • Every instrument the script uses is marked busy for its duration, so its own console is locked out exactly as a single-instrument script does.
  • Results are rows, saved as CSV — the deliverable of a sweep is a table to plot.
  • The run log and the results each have their own pair of buttons, under the pane they act on: Save Log / Clear Log below the log, Save CSV / Clear Results below the table. They are independent — clearing the table must not discard the log, which is the record of how the table was filled.
  • The results area carries both a table and a plot of the same rows. The table is the record and is what gets exported; the plot is how the measurement is read, because forty rows of numbers are not a frequency response and the shape is the answer. It redraws per recorded row, so a sweep going somewhere wrong is visible on the tenth point rather than after the fortieth.
  • Which column goes on which axis is chosen, not guessed — a script may record anything in any order. The default is the first column across and everything else up, which is right for a sweep and means a plot appears rather than needing assembly. Log axes are offered per axis and disabled when any value is zero or negative: a log axis through zero has no meaning, and quietly dropping those points would hide data.
  • A reading that failed to parse drops that point from that curve only. Plotting it as zero would invent a measurement; dropping the row would delete another instrument's reading.
  • The bundled examples carry a representative *IDN? per declared model, so each line can be checked against the catalog of the instrument it is addressed to. Same rule as §10, and it earns its keep: a Siglent generator's frequency is C1:BSWV FRQ, not FREQ.
  • One belongs to the bench, not to a console: Tools ▸ Multi-Instrument Scripts…, or the button beside Timeout on the connect row. Modeless and single-instance, so instruments can be connected in the main window while it is open.

9b. The editor teaches the language

The language is this application's own. The commands inside it are SCPI, which is a real standard; everything around them — DEVICE, WITH, FOR…POINTS…LOG, RECORD, ->, $name — was invented here and exists nowhere else. Nobody arrives already knowing it, so the editor teaches it rather than assuming it. Both editors share one control.

  • Colour marks the shape of a line: comments, keywords, instrument aliases, $values, numbers. The SCPI itself is left plain — it is the content, not the frame — and a page of seven colours reads worse than a page of three.
  • A leading word: is coloured as an instrument only when a DEVICE line declared it. C1:BSWV and gen: are the same shape; claiming the wrong one is a lie the eye believes.
  • Snippets — a dropdown naming every construct with a description, and the whole language on one page behind it. Choosing one writes it in with its blanks «selected»; Tab steps to the next. Typing a trigger word and pressing Tab does the same thing.
  • Completion offers keywords, snippets, the aliases this script declared, the values it captured, and the documented commands of the instruments in play. Ctrl+Space forces it. After a $ only captured values are offered, because $ has exactly one meaning.
  • The editor's word list and the runner's are checked against each other by test. A keyword the editor offers and the runner rejects teaches the wrong thing, confidently.

10. Command discovery and curated references

There is no universal way to enumerate an instrument's command set — the programming manual is ground truth.

  1. Try SCPI-99's SYSTem:HELP:HEADers?. A reply counts as genuine only if it has at least three lines and most of them look like SCPI headers; a stray error line or a bare 0 is a failure, not a command list.
  2. On failure, open the curated catalog for the instrument's family, if one is bundled.
  3. If neither is available, say so plainly and point at the manual.

Catalogs are JSON, embedded as commands.<family>.json, each entry carrying category, syntax, description, optional example, whether it is a query, and how far it is trusted. Each catalog records the guide it was transcribed from.

36 catalogs, 23,978 entries, of which 518 carry a bench tick and 2,778 a cross-check.

No description stops mid-sentence. A guide wraps a description at the margin, and the extractors used to stop at the line it started on whenever the next line began with something command-shaped — so "…the external generator selected with" lost its second half, and the full stop emit adds to every description made the cut look deliberate. Six readers had the fault, each in its own shape: breaking on a command-shaped continuation, never joining at all, stopping at the blank line inside a wrapped sentence, and stopping at the label a wrap runs into in a two-column layout.

All 129 were repaired from the guides they came from — which meant fetching thirteen guides that were missing, the last of them by hand because its URL serves a landing page rather than the file. CatalogDescriptionTests keeps the count at zero per family; a family is listed there only when its guide cannot be had and the missing words would have to be invented, which is now true of none of them.

A further 307 were finished the same way, and none of them looked broken. A sentence cut after "…the next data after the last fetched" or "…an event on a selected trigger" ends on an ordinary noun, so no rule catches it — they were found by re-reading every guide with the mended readers and keeping only text that continues what shipped. 54 command names quoted in prose lost a stray space at the same time ("CALC:LIM:CONT: OFFS" was never a command).

A description ending in a colon now keeps its list. The guides put the alternatives a parameter accepts in a table under the sentence that introduces them, so "sets the sort mode for the search for maxima:" answered everything except what the modes are. 60 of the 92 have been filled in — "…: X — the maxima are sorted … according to increasing X values; Y — …" — and the rest are tables too wide or too deep to read as a sentence. Only a description ending in a colon is extended, and only when what precedes it reads as a sentence rather than a heading: "Impedance:" over a list of related commands is a label, and treating it as a promise gave three FPC entries a description made entirely of cross-references.

One of the 49 was never damage. English stands a preposition at the end of a sentence after certain verbs, and "…the type of information that a test report consists of." finishes on the same word a severed sentence would. The check now allows that handful of verbs and nothing looser, because a cut that goes uncounted is the failure it exists to catch.

Family Entries Bench ✓ Cross-checked • Source
R&S FSW analyzer 2358 0 111 R&S FSW User Manual (1173.9411.02 v56), chapter 13
R&S FSU analyzer 975 0 91 R&S FSU Operating Manual (1313.9646.12-02), chapter 6
R&S FSP analyzer 1044 0 92 R&S FSP Operating Manual (1164.4556.12-02), chapter 6
R&S FSQ analyzer 997 0 91 R&S FSQ Operating Manual (1313.9681.12-02), chapter 6
R&S FSL analyzer 2256 0 119 R&S FSL Operating Manual (1300.2519.12-12)
Tektronix scope 2031 0 186 Tektronix MDO4000C/MDO4000B/MSO4000B/DPO4000B/MDO3000 Programmer Manual
Keysight scope 1793 0 142 Keysight InfiniiVision 3000T X-Series Programmer's Guide (9018-07265)
R&S scope 1449 0 85 R&S RTB2000 User Manual (1333.1611.02 v09), plus the RTM3000 and RTA4000 manuals
R&S FSV analyzer 1279 0 109 R&S FSVA/FSV Operating Manual (1307.9331.12-17)
Oscilloscope 1200 409 147 Rigol MSO2000A/DS2000A Programming Guide (Feb 2016)
R&S FSIQ analyzer 1051 0 0 R&S FSIQ Operating Manual (1119.5063.12), chapter 6 — first extraction, no adoption pass
Siglent scope 859 0 120 Siglent SDS Series Programming Guide (EN11D) + SDS3000X HD (EN11F)
Rigol spectrum analyzer 586 0 88 Rigol DSA800 Series Programming Guide (Aug. 2016)
R&S spectrum analyzer 537 0 99 R&S FPC Spectrum Analyzer User Manual (1178.4130.02 ─ 08)
R&S power supply 434 0 139 R&S NGL200/NGM200 User Manual, plus the NGE100 and HMP guides
Waveform generator 430 0 89 Rigol DG1000Z Programming Guide
Keysight multimeter 390 0 100 Keysight Truevolt Series Operating and Service Guide
GW Instek GDS-1000B scope 383 0 69 GW Instek GDS-1000B Series Programming Manual (v1.10)
Keithley multimeter 376 0 78 Keithley DMM6500 Reference Manual (DMM6500-901-01 Rev. A)
Siglent generator 343 27 42 Siglent SDG Series Programming Guide (PG02_E05B)
Chroma electronic load 339 0 31 Chroma 63200A Series Operation & Programming Manual (Oct 2024)
Power supply 309 0 65 Rigol DP800 Programming Guide (Dec 2015)
Keithley SMU 293 0 74 Keithley Model 2450 SourceMeter Reference Manual (2450-901-01 Rev. E)
Electronic load 292 0 29 Siglent SDL1000X Programming Guide (E02B)
Chroma modular load 286 0 30 Chroma 63600 Series Operation & Programming Manual (V2.2)
Multimeter 206 82 20 Siglent SDM Series Programming Guide (EN02A)
Spectrum analyzer 202 0 55 Siglent SSA3000X Programming Guide (PG0703X-E03D)
Rigol multimeter 186 0 28 Rigol Programming Guide for DM3058/DM3058E (Jan. 2015)
Keysight power supply 171 0 61 Keysight E36300 Series Programming Guide (9018-04577)
GW Instek scope 169 0 36 GW Instek GDS-2000 Series Programming Manual
B&K triple-output supply 144 0 58 B&K Precision 9130B Series Programming Manual (V051415) — 22 entries flagged misprinted
Rigol electronic load 144 0 59 Rigol DL3000 Series Programming Guide (Apr. 2019)
B&K electronic load 136 0 64 B&K Precision 8600 Series Programming Manual
Fluke multimeter 125 0 77 Fluke 8845A/8846A Programmers Manual (Sept 2006), hand-transcribed
Chroma power supply 121 0 51 Chroma 62000L Series User Manual, Remote Control Reference
B&K power supply 84 0 43 B&K Precision 9200B Series User Manual, chapter 5 — 16 entries flagged misprinted

The counts are generated from the catalogs rather than kept by hand — several rows had drifted by one or two before that was noticed, which is exactly how much a hand-maintained table can be wrong without anyone seeing it. node tools/catalog-table.js regenerates them and CatalogTableTests fails if they are stale, because for a while "generated" meant generated once: the numbers had been read out of the catalogs by hand and every edit since drifted away from them.

The two marks mean different things, and only one of them needs the bench. A says the command was sent to the real instrument and answered. A says the same command was found in an independent open-source driver — pymeasure, python-ivi, QCoDeS, tm_devices, OriginalCircuit — which is weaker evidence, and free: it needs no hardware, only the driver sources, and tools/scpi-extract/README.md gives the clone lines that produce the corpora. Those corpora are not committed, and while they were absent every catalog built without them silently gained no marks at all. Re-running the check against 7,581 driver commands took the cross-checked count from 1,460 to 2,778, twenty catalogs gaining marks and twelve of them going from none. The bench count did not move and cannot for now: all three instruments here are unreachable, so is stuck at 518 until they are back.

One catalog corrects its guide. The Chroma manual prints VOLTaget:PROTection and PROTecton:TRIPped? a handful of times while spelling both correctly everywhere else; transcribing the slip faithfully would put commands in the catalog that the instrument rejects, so the build config lists the corrections explicitly.

Two catalogs do the opposite, deliberately. Both B&K supply guides misprint commands in places where nothing else in the document settles what was meant — a dropped letter in a query whose set form is spelled correctly, an index and a described entry that disagree, a subsystem written two ways, a node spelled three ways across two manuals from the same vendor. Correcting those would be guessing, so the entry carries the syntax exactly as printed and a guideMisprint note saying what the guide prints, why it looks wrong, and what to try instead. The library shows those entries with a . 22 entries in the 9130B catalog carry one and 16 in the 9200B.

The treatments are not in tension: Chroma's slips are settled by the same guide, which spells the commands correctly elsewhere, and most of B&K's are not settled by anything short of the instrument. Where the guide answers the question, follow it — the 9200B manual states in its own §5.1 that a command is separated from its parameter by a space, so the three entries it prints without one are written with the space, and flagged as corrected. Where the guide does not answer, say so rather than inventing an answer. Nothing may be corrected silently under either.

A wrapped line is not a command. Guides break a long name across two lines — SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:FLEXray:HEADER: then PAYLength? — and a line-based extractor reads the continuation as a command of its own. The catalog then ships PAYLength?, which no instrument answers, and loses the query form of the command it came from. Eight such entries were found in the Tektronix catalog, plus a ninth where the fragment had overwritten a real root command's description. The truncated-line check does not catch these: a bare PAYLength? has balanced brackets and a letter in its header, so it is well-formed in every way except being real. A separate invariant now looks for a single-node entry that repeats, word for word, the description of a longer entry ending in the same mnemonic.

A transcription can also be wrong where the guide is right, and that failure is quieter: the catalog still loads and the library still lists the entry. The 9200B's first, machine-assisted pass paired eleven entries with a neighbour's description — VOLTage:LIMIt explained as a query of the OVP state — dropped both STEP query forms and the whole VOLTage:TRIGgered pair, and carried five commands only in the abbreviated spelling their syntax lines use, which stops the full spelling in the guide's own headings from matching (ScpiSyntax derives the short form from the template, so a template that is already short admits nothing else). Tests now pin each class.

One catalog is hand-transcribed rather than extracted: the Fluke 8845A/8846A guide is a set of indented tree tables whose description column pdftotext shifts by a row in places — Table 12 pairs :DIODe? with "Make a frequency measurement" — so an automatic pass produced confidently wrong entries and invented paths like :CURRent:VOLTage. Those were discarded and the catalog read off the guide by eye instead. 77 of its 125 entries are corroborated by an independent driver.

One catalog carries commands the guide does not print verbatim: the R&S manual states in its SCPI conventions that "a query is defined for each setting command unless explicitly specified otherwise", and spells the exception out per entry in a Usage field (Query only, Setting only, Event). Query forms are generated from that rule where Usage permits — applying a documented convention, not guessing.

Two levels of confidence, both weaker than the guide is authoritative:

  • BenchVerified — the command was sent to the real instrument here and answered.
  • CrossChecked — the same header was also found in an independent open-source instrument driver (pymeasure, python-ivi, QCoDeS, tm_devices, OriginalCircuit). This catches a command transcribed correctly from a guide that the hardware never honoured.

How much of this is right

Everything above describes how a catalog is built and what is checked. None of it says how much of the result is correct, and for a long time nothing did. Every count this project produced answered "how many entries match a pattern someone thought to look for" — which measures the search, not the data. Fixing one such pattern reliably exposed another, and there was no way to tell a long tail from an endless one.

So it was measured, on 15 August 2026. tools/sample-catalog-accuracy.js draws entries uniformly from all 23,978 — including the ones whose guide is not in this checkout, which are judged unverifiable rather than dropped, since dropping them would quietly measure the best-documented half. The draw is seeded, so the same sample can be redrawn and rechecked. 100 entries, each read against its guide:

7 defective. A 7% error rate, 95% Wilson interval 3.4% to 13.7% — between 800 and 3,300 entries across the catalogs, most likely about 1,700.

83 of the 100 had guide text to compare against; the other 17 were judged on internal evidence, so if any of those are wrong the true rate is higher, not lower.

The seven were: two with a margin label spliced into the sentence, one describing a different command outright, one opening on a stray digit, one promising three values and giving one, one trailing off mid-clause, one missing a space.

Two of the seven turned out to belong to classes nobody had counted — 323 Rigol descriptions opening on the tail of the guide's "Description 1", and 89 Keysight descriptions carrying a left-margin label inline. Both were repaired the same day (421 entries with the strays they exposed) and are now held by No_description_carries_the_guide_page_furniture. Neither had ever surfaced from reading the catalogs directly; both fell out of a hundred random entries within an hour.

The rate above is therefore stale by construction — it was measured before that repair, and those 421 entries are about a quarter of what it projected. It is left as measured rather than adjusted by arithmetic. Redraw the sample with a new seed before quoting a number.

The distribution matters more than the headline. The R&S catalogs, roughly half of all entries, gave 1 defect in 28 sampled; the Rigol and Keysight scopes gave 4 in 12. Remaining work is concentrated, not spread.

What is known to be left, as distinct from what the sample estimates:

347 descriptions cut at a character count that tools/finish-capped-descriptions.js declined to finish — 235 where the guide does not carry the words they stop on, 66 with no heading found, 44 that were already two descriptions merged before the cut, 2 where the guide says no more. The 44 are wrong rather than short.
47 descriptions shared between commands with different roots. Sampled and read as correct — a guide really does give *LRN? and SET? one sentence — but not audited one by one.
13 catalogs never audited for mispaired descriptions. Every one audited so far (Tektronix, the Keysight DMM, the R&S scope) found some.
4 B&K 9130B entries whose guide misspells INSTrument as INSTument in both its index and its body. Only a 9130B can say whether the firmware accepts the misspelling, and there is none on this bench.

None of these blocks anything. They are recorded so the next person does not have to rediscover them, and so that "what else is left" has an answer that is not a search.

Never invent SCPI. If a command cannot be transcribed from the guide, leave it out. This is enforced, not merely stated: CatalogCoverageTests fails the build if any quick-command button, live-readout query or bundled script line is not an instance of a syntax template in that family's catalog. ScpiSyntax does the matching, and understands short/long mnemonic forms, optional nodes, bracketed roots, channel suffixes and inline alternation — so :OUTPut2 OFF is recognised as a use of :OUTPut[<n>][:STATe] {ON|1|OFF|0}.

The guides themselves:

Two more turned up later. The FSIQ's has since been catalogued — a first extraction, its caveats recorded in the catalog's own source field — and the FSE's cannot be: the copy is Volume 1, and chapter 6, the command reference, is in Volume 2:

And one was found and cannot be used, which is worth recording so nobody hunts it twice:

  • Chroma 63800 Series Operation & Programming Manual, Version 1.1, April 2009 — https://web.archive.org/web/20250805042045id_/https://www.transcat.com/media/pdf/Chroma-63800-Series-User-Manual.pdf

    The live Transcat copy now answers 403 and no other mirror is known; the Wayback capture of 5 August 2025 serves the original bytes intact. It is the right document, and it is a scan: 44 MB of page images with no text layer, from which pdftotext recovers 550 bytes — the cover, and nothing else. Extraction has nothing to read. Making it usable means OCR, and OCR of a 2009 scan produces command spellings that would have to be reviewed character by character against the page images before any of it could be trusted, which is a different undertaking from transcription and is not planned. The 638 AC loads therefore stay routed to Generic (§8, row 16) — by document, not by oversight.

pdftotext -layout makes them far cheaper to work from than reading page images. The extraction pipeline built on that is in tools/scpi-extract, and each catalog's own source field records which guide it came from, what that guide covers, and what was deliberately left out.

How far a catalog can be rebuilt from its guide. Each config records its parse recipe — manual, reader style, any flags — and node rebuild.js --all runs the lot and reports where each one stands. Before that the recipe was written down nowhere: a config named the extracted files it was built from, but those live in parsed/, which is regenerated and not committed, so the step that produces them survived only in whoever last ran it. Eleven catalogs were listed as unrebuildable for having no config, which was true and beside the point — none of the other twenty-four could be rebuilt either.

As measured on 14 August 2026, after fetching twelve guides that were missing: 1 rebuilds byte-identically, 21 rebuild but differ, and 2 cannot be attempted — the Keysight scope because its guide no longer serves as a file, and Fluke because it is curated by hand and extracts nothing. Before those guides were fetched only twelve of the twenty-four could be attempted at all.

"Differs" is the expected state, not a defect to close. A catalog is extraction plus curation, and the curation is load-bearing: query forms the guide states in prose rather than prints, entries transcribed by hand where the layout defeats the reader, and junk removed after reading the page it came from. Re-emitting throws all of that away and keeps only what a parser can see.

Measured, rather than assumed. A wholesale re-emit today would gain 153 entries and lose 129, and the losses are real commands: *CAL? and *IST? from the R&S scope, ALIas? and CH<x>? from the Tektronix, :ERRor? and :VERsion? from the B&K load, the whole :HARDcopy tree from the GW Instek, READ? and both MEASure queries from the R&S supply. Meanwhile the "gains" include *PSC {OFF|ON|NR1>} with a bracket missing, CH<x>:PRObe:MODel<String> with its space eaten, and — in the supply catalog — RUN and RUN?, the parameter values deleted by hand earlier the same day.

So the pipeline is for building a catalog and for checking what a guide contains, not for regenerating one that people have since worked on. node rebuild.js --all reports where each stands; a catalog that has drifted is telling you curation happened, and the way to harvest a genuine addition from a re-read is to adopt it deliberately, entry by entry, not to overwrite the file. A missing dump is likewise not usually a defect: manuals/ is ignored for the same reason datasheets/ is, and every guide is named and linked in the list above.

11. Capture

Screen. For instruments with a known screen-dump command, query it as an IEEE 488.2 binary block and show the image in a viewer that can save it. The transfer is given headroom over the user's timeout (at least 15 s) — a full-screen BMP is around 1 MB — and the timeout is restored afterwards. Most vendors take the image format as a parameter of the query (:PRINt? BMP); Tektronix takes it as separate state, so a profile may also carry setup commands to send first.

Offered for Rigol (:DISPlay:DATA?), Keysight (:DISPlay:DATA? PNG,COLor), Tektronix (SAVe:IMAGe:FILEFormat PNG then HARDCopy STARt), R&S (HCOPy:DATA?) and Siglent (:PRINt? BMP).

Waveform. Read channel 1, decode, plot, and offer CSV. At least 10 s of headroom.

Five dialects, because no two vendors agree on the command tree, the way the scaling is described, or the arithmetic that turns a stored sample into volts. Each is transcribed from that vendor's guide and named in WaveformDialect; the sequences live in Core/Capture/WaveformReader so they can be exercised against a fake client.

Dialect Reads Scaling
Rigol :WAVeform:DATA? in BYTE 10-field text preamble; v = (raw - yref - yorig) * yinc
Keysight :WAVeform:DATA? in BYTE the same 10 fields; v = ((raw - yref) * yinc) + yorig
Tektronix CURVe? WFMOutpre field by field; v = ((raw - YOFf) * YMUlt) + YZEro
RohdeAscii CHANnel<m>:DATA? in ASCII none — the instrument returns volts
Siglent :WAVeform:DATA? binary descriptor; v = code * (vdiv / code_per_div) - voffset

Rigol and Keysight are the trap. Rigol modelled its command set on Agilent's, so the preamble is field-for-field identical and the decoder looks reusable. It is not: the two formulae agree exactly on a trace centred at zero and differ by the offset everywhere else, and both draw something that looks like a measurement. A test exists whose only job is to fail if the two ever return the same number.

A preamble that cannot carry the scaling — fewer than 10 fields, a descriptor too short, zero codes per division — is an error, reported as one.

Where capture is refused. A GW Instek GDS-2000 offers neither, and in both cases because its manual documents no way rather than because the work is outstanding. :COPY writes the screen to a flash disk or a printer, never back over the wire. :ACQuire<X>:MEMory? does return samples, and the framing is precise — eight header bytes, the sample interval as a little-endian float, then two bytes a point MSB first — but the manual never states how a stored code becomes a voltage. That needs a constant nobody wrote down. A trace drawn against a guessed vertical scale is wrong by an unknown factor and looks entirely convincing, which is the same reason §10 refuses to invent SCPI.

Live readout. Instruments that answer one measurement per query — a multimeter, a power supply or an electronic load, but not a scope, which already plots against time itself — offer a window that polls one function on a user-set interval (100 ms to 60 s, default 1000 ms) and plots the values against time, with the current value shown large. The profile decides which functions are offered (ReadoutFunctions); an instrument with none does not get the window. A supply offers per-channel volts, amps and watts; a load adds resistance.

  • Polling holds the link, so it marks the session busy exactly as a running script does and the console locks itself out (§7).
  • The series is bounded — a meter polled all afternoon must not grow without limit — and says so when it is dropping the oldest readings.
  • Changing the measured function clears the plot: two quantities on one axis mean nothing.
  • Readings are parsed invariantly. On a machine whose locale uses a decimal comma, culture-sensitive parsing would read 1.25 as 125 and silently corrupt every value.

11a. Browsing the catalogs

Tools ▸ Command Library opens every bundled catalog for reading, whether or not anything is connected. Manufacturers on the left, that maker's instruments beneath them, the selected catalog's commands on the right, and one filter box matching maker, model, instrument or command text.

Each catalog names the guide it was transcribed from, and the panel offers to open the vendor's page for it. It will also open a local copy, if the user has pointed the app at a folder of downloaded PDFs — a per-user setting, off until set. The guides themselves are never bundled: they are the vendors' copyright, they are revised independently of this app, and a stale bundled copy is wrong in a way that is hard to notice.

This window shows only the curated catalogs. Anything a model extracted (§11b) is the user's own material and belongs to the instrument it was extracted for, not to a library of things the app vouches for.

11b. AI datasheet extraction

For an instrument no bundled catalog covers, the user may connect their own AI provider and have it read a datasheet — PDF, .docx or plain text. This is optional, off until configured, and there is no built-in key: the user brings their own.

The key. Stored encrypted with DPAPI scoped to the current Windows user, so a settings.json that is copied, synced or backed up elsewhere carries nothing usable. It is never written in the clear and never logged. A settings file from another machine decrypts to nothing, which is treated as "no key set" rather than as an error.

What is sent. Only the document the user chose, to the provider the user chose. Where a provider accepts a PDF directly that is preferred, because flattening a two-column programming guide to text interleaves its columns — which is how a command ends up with its neighbour's description. Providers without file support get text extracted locally instead, and the checkbox that controls this explains the cost either way.

Refused before sending, not after. Provider limits are checked against the file first — Gemini 50 MB and 1000 pages, Anthropic 600 pages and 32 MB of request payload, which base64 inflates by a third, so a 25 MB PDF fails a 32 MB cap. An upload that cannot succeed is refused with the reason and the actual numbers, not with the provider's error an upload later.

Nothing is trusted on the way out. Every candidate command is checked for the shape of SCPI before the user sees it (ScpiSyntax.IsValidTemplate), which is what rejects a model returning a sentence where a command was asked for. What survives is shown for review and saved only if the user accepts it.

Kept apart. Extracted commands never join a curated catalog. They are stored per instrument, keyed on the model rather than the address — an instrument on DHCP moves, and its commands should follow the instrument, not the lease — and marked everywhere they appear. A catalog in this repository means someone transcribed a vendor's guide (§10); that guarantee is worth keeping distinct from a machine's best guess, however good.

11c. AI script writing

Both editors — the per-instrument one (§9) and the sequence one (§9a) — have a Write with AI… button. The user describes a measurement in plain English and gets a script back. It uses the same connection and the same stored key as §11b, and is equally optional.

What is sent. The request, the command catalogs of the instruments involved, the script language, and — when the user leaves the boxes ticked — the script currently in the editor and the tail of the last run's output. The output is what makes "it failed with -113, fix it" answerable: the failure exists only in that log.

Given the catalog, not asked to remember one. A model writing SCPI unaided reaches for whatever dialect it has seen most, which is how :SOURce1:FREQuency 1000 gets sent to a Siglent generator that has never heard of it. So the commands go in the prompt and the model is told to use nothing else. A large catalog is trimmed to the commands the request is about — the R&S FSL alone is 2,270, and a model given all of them chooses worse than one given the plausible few, quite apart from the cost.

Checked on the way back. Every command header in the returned script is matched against the catalog of the instrument that line addresses, which for a sequence means a different catalog per alias. Anything unmatched is listed under the draft. The check is a header check: a documented header carrying a wrong argument passes it, and the window says so rather than letting silence read as "verified".

Never run, never saved. What comes back is a draft in a preview pane. It reaches the editor when the user presses Use, and runs when the user presses Run — two deliberate acts, because these scripts turn on outputs and apply voltages to real circuits. Nothing generated is written to a catalog; §11b's separation holds here too.

12. Files and formats

Output Format
Scan results CSV, RFC 4180, header IP Address,Port,Protocol,Identity, CRLF, fields quoted only when needed
Console log Plain text, exactly as displayed
Waveform CSV, header Time (s),Voltage (V), g9 invariant-culture numbers
Settings JSON at %AppData%\LabEquipmentController\settings.json

Settings persist the last interface address, port list, communication timeout, and the window size and maximized state. A missing, empty or corrupt settings file must yield defaults — persisted preferences must never block startup, and a failed write must never block shutdown.

Open tabs are not persisted: the instruments are on DHCP and their addresses move, so reconnecting on launch would be guesswork (and risks §13).

13. Transports and instrument-specific behaviour

IInstrumentClient is the transport abstraction — connect, send, query, query-binary, return-to-local, close — with three implementations. The UI must not care which is beneath it, and neither must the sessions, the script runners, the capture path or the catalogs.

Raw socket. TCP; commands terminated with a newline; replies read to the terminator.

VXI-11. A hand-written ONC-RPC client: portmapper GETPORT to find the core channel, then create_link, device_write, device_read, device_clear, device_local, destroy_link. XDR encoding, big-endian, 4-byte record marking.

Serial (RS-232). A serial port opened at the line settings the address named (§5), System.IO.Ports underneath, so COM3 on Windows and /dev/ttyUSB0 or /dev/cu.usbserial-* elsewhere all work — including the USB-to-serial adapter most benches actually use. DTR is asserted on open, and RTS too unless hardware flow control is driving it; a good many instruments and nearly every adapter want to see them before they will talk. Four things differ from the LAN transports and nothing else does:

  • The line settings have to be right first. There is nothing on the wire to negotiate them with, and a wrong baud rate does not fail — it answers with rubbish.
  • The terminator is not always LF. A good many RS-232 instruments want CRLF and treat a bare LF as no command at all, which looks exactly like a dead port. Replies are read to the line feed either way, with carriage returns discarded.
  • Nothing happens when the port closes. A raw socket returns the front panel to local by being closed; RS-232 has no such signal, so the request is made in band with SYSTem:LOCal — SCPI-99's own answer for an interface with no bus-level Go To Local, and documented in 12 of the 36 catalogs. Best-effort: on an instrument that does not have it, it costs one entry in an error queue that is about to be disconnected from.
  • A binary read has no total deadline, only a gap-between-bytes one, which the driver enforces. That is the better question to ask of a transport this slow: 100 KB of screenshot at 9600 baud is a minute and three quarters of perfectly healthy transfer, and no fixed total that allowed for it would still catch an instrument that had stopped talking. A socket uses a total deadline only because an async read on a NetworkStream ignores its own read timeout.

Discovery extends to it only as far as the physics allows (§4, §17). Ports are listed without being opened — the CLI's lec ports, SerialPorts.Names() — and that listing is what arriving at the card gives you. Opening them is a separate act on a separate button: SerialScanner.ScanAsync() tries the baud rates it was given against the ports it was given, and nothing else. A subnet sweep can ask every address the same harmless question; a serial port cannot be asked anything until the settings already match, and what is on the other end may be a printer or a modem.

IEEE 488.2 blocks. #<n><length><data> definite-length blocks are parsed and the payload returned with the header and any trailing newline stripped. A response that is not a block is returned whole. The framing — writing a command, reading a line, reading a block — is one implementation (ScpiFraming), shared by the socket and the serial port, because SCPI over RS-232 is line-based in exactly the way SCPI over a raw socket is. It is where the careful parts live and they are worth having once: a block is read by its declared length rather than until a pause, since binary data contains newlines and arrives in pieces.

Hard-won facts the software must respect

Rigol DS2202 oscilloscope

  • Its raw socket (port 5555) is broken: every query returns the previous query's answer, permanently, across reconnects. A known DS2000 firmware bug. Hence: never send *IDN? on a raw probe, and prefer VXI-11 whenever the instrument offers both.
  • VXI-11 framing quirk: it sends empty, non-END device_read packets before its data and part-way through a large transfer, while it produces more. A read loop must continue until the END flag; it must not stop on the first empty read, nor on a count of consecutive empty reads. Bound the wait on elapsed idle time instead. (The Siglents never do this, which masked the bug for a long time.)
    • Measured: giving up after 10 empties truncated a screen dump at exactly 64 KB — Block claims 1152054 bytes but only 65525 remain. Waiting on the clock instead returns the full 1,152,054 bytes in ~10 s. A whole screen dump is ~1.15 MB, so the read must tolerate a transfer running for many seconds.
  • Issue device_clear on connect to flush a stale output queue.
  • A poisoned output queue survives one device_clear. If a large binary transfer is aborted (app killed mid-:DISPlay:DATA?), the leftover payload is returned as the answer to the next query — *IDN? comes back as thousands of BMP padding bytes, or as a bare #9000001400 block header. Reconnecting once more clears it; the instrument does not need a power cycle for this. Do not confuse it with the wedge below.
  • One TCP session only. Rapid or parallel connections wedge the firmware — no ping, ARP entry goes Incomplete, front panel dead, though the trace keeps running. Only a power-cycle recovers it. This includes automated test loops: repeated connect/disconnect cycles from a test harness will wedge it just as surely as a user double-clicking. Space them out, and stop at the first sign of trouble.

Siglent SDG2042X generator

  • VXI-11 only — no raw SCPI socket at all. Port 23 is a BusyBox root shell, not SCPI.
  • Uses Siglent's own dialect: C1:BSWV WVTP,SINE, C1:OUTP ON. Not standard SCPI.

Siglent SDM3065X multimeter

  • Standard SCPI (MEASure:VOLTage:DC?) despite sharing a maker with the SDG. This is why family classification checks multimeters before generators (§8).

Instruments are on DHCP and their addresses move. Match on the *IDN? model; never hardcode an IP.

14. UI conventions

Rules for building a UI in WinForms at all. Which controls exist, in what order, with what words on them — for this build and for the web one — is UI-SPEC.md.

  • Every control gets a hover tooltip describing what it does. The labels alone don't explain the controls. This is a standing requirement, not a one-off — anything added gets an entry in the owning window's SetTooltips().
  • Rows share one height. WinForms gives ComboBox, TextBox and Button different heights under AutoScaleMode.Font, so heights are normalised in code. A button added to an existing row must be pinned there too, or it visibly stands taller.
  • Never hardcode pixel sizes for controls created at runtime — they miss the form's auto-scale pass. Copy metrics from an existing designer control, or use dock/auto-size layout and normalise afterwards. This includes the height of a button bar: fixed values clipped the buttons in both capture windows (one had already been nudged 38 → 54 and clipped again), so a bar either auto-sizes or takes its height from button.PreferredSize.Height plus its own padding, in OnLoad.
  • A docked Label with AutoSize left on ignores a Height set in code, and clips its own text. Turn AutoSize off for any label whose size the layout controls.
  • Every button in the app is built and sized through ButtonStyle — one padding, one margin, one nominal 16-logical-pixel glyph size, one minimum width, and one height derived from the font rather than from whichever control a given window happened to measure. Windows that each measured their own tallest control drifted apart: the main window shipped 23, 20 and 23 px buttons with 13, 12 and 15 px glyphs and one lone FlatStyle.Flat, while the console used three heights in one control and the script editor took its combo box's. Do not hand-set a button's Padding, Height or glyph size; add to ButtonStyle instead.
    • Glyphs are corrected optically, not numerically (ButtonStyle.Optical). The bundled artwork does not share a common margin: the play triangle and stop square are solid shapes drawn edge to edge in their canvas, while the file and magnifier glyphs are thin outlines inset from theirs. At one nominal size the solid ones read as roughly twice the weight — on the console's Run/Stop/Single row they looked like blocks. New artwork that fills its canvas needs an entry in that table; judge it from a glyph sheet rendered at the real 16 px, not from the numbers.
    • A NumericUpDown cannot be grown. UpDownBase.SetBoundsCore clamps it to a font-driven height, MinimumSize and all. It — and any label — is centred on the row with ButtonStyle.CentreInRow instead, because a FlowLayoutPanel hangs children from the top of the row.
    • A DropDownList ComboBox can, but only owner-drawn, through ItemHeight (ButtonStyle.MatchHeight). Paint it with ButtonStyle.DrawComboItem: WinForms hands the handler the selection colours for the closed box whenever the combo has focus, so drawing what it gives you puts a blue bar across the control as soon as it is tabbed into.
    • A single-line TextBox cannot either. Anchor it Left|Right with no vertical anchor and a TableLayoutPanel will centre it in the cell rather than pin it to the top.
  • Adding a glyph to a button widens it, so a wrapping button bar that fitted on one row may quietly fold onto two. Check the window at its default size after adding one, and pay for the row in horizontal padding before paying for it in window width.
  • A button that toggles its own label (Start/Stop) pins its width to the wider of the two states, measured through PreferredSize. Left to auto-size it changes width on every press and shunts the buttons beside it sideways.
  • Glyphs with no bundled artwork are drawn at runtime (AppIcons.Drawn). They render at roughly 16 device pixels, so keep them to one idea — an early detach icon drew a window and an arrow and an arrowhead, and read as a smudge.
  • Long, variable text (instrument identities) is clipped with an ellipsis on one line, not wrapped, so panel heights don't shift with the window width. The full text stays available in a tooltip.
  • The console log is a dark, fixed-pitch surface; colour carries meaning (§7).
  • Cosmetic failures — a missing glyph, an unreadable icon — must never break or block the UI.
  • MainForm.Designer.cs is hand-edited. It still opens in the Visual Studio designer, but most of the layout arithmetic lives in MainForm.cs, and the designer will not round-trip the comments. Edit it as source.
  • Don't assume a fixed display scale. This bench has been seen at both 120 and 168 DeviceDpi within one session (a Hyper-V guest's DPI follows its viewer window), and the form's MinimumSize scales with it — a window asked for below that minimum is silently clamped, which reads as a cramped layout rather than as a sizing failure.

15. Non-functional requirements

  • The UI thread never blocks. Scans, connections and transfers are asynchronous; long synchronous ramp-up is pushed to a worker. The window stays responsive and Stop is handled promptly during a full-subnet scan.
  • Teardown is best effort and bounded. An instrument that has been switched off must not stop a session closing or hang application exit.
  • Failures are reported, not swallowed silently, and never fatal. A failed command, a corrupt catalog, an unreachable host: each degrades locally.
  • No telemetry, no network access beyond the instruments on the selected subnet.

16. Quality bar

There is no CI. The standard applied to changes:

  1. Build and dotnet test for anything touching Core. All logic that can live UI-free does, so that it can be tested against a fake instrument client.
  2. Screenshot the running app for UI changes — a UI change is not verified from code alone.
  3. State plainly what was not verified. Most features have never run against the real instruments (they are often powered off, or the bench network is unreachable). That is always said rather than glossed over.

Do not drive the physical mouse (SetCursorPos) while testing — the user is normally working on the same machine.

Traps that have cost real time, all worth knowing before writing another harness:

  • Drive the app from inside Application.Run, never Show() + an Application.DoEvents pump. Without the real message loop the WinForms SynchronizationContext does not govern await continuations, so the code after await client.ConnectAsync() in ConnectSelectedAsync resumes on a thread-pool thread — and the InstrumentConsole created there belongs to that thread. Reparenting it later throws "Controls created on one thread cannot be parented to a control on a different thread", which surfaces from Controls.Remove as a bare ERROR_INVALID_WINDOW_HANDLE. This masqueraded as a detach bug across several bench runs and is entirely an artefact of the harness. Installing the context by hand is not enough; run the script from Form.Shown with real awaits. Check it directly: a console's InvokeRequired must be false from the UI thread.
  • A UI harness that pumps Application.DoEvents() deadlocks on any MessageBox. When an operation fails, the app shows a modal dialog; the dialog's own message loop starts inside the DoEvents call and never returns. Every later reflection-driven call then runs in a modal-blocked, re-entrant state, where Controls.Remove fails with Error creating window handle and screenshots capture stale paint. The symptoms look like deep WinForms bugs and are nothing of the sort. Detect it: enumerate thread windows for class #32770, and dismiss from a separate thread — the UI thread is blocked.
  • A DPI-unaware harness gets virtualized window coordinates. GetWindowRect from a process that has not called SetProcessDpiAwarenessContext reports a 1563 px window as 909 on this bench's 168 DPI display, so a capture sized from it comes out clipped at the right edge — with exactly the newest controls missing. Declare per-monitor-v2 awareness first.
  • GetWindowText is not a reliable read-back across processes. It returns the window's own text, not the contents of another process's edit control — it happened to answer for the address box and came back empty for the console's command box, which did contain the text. Verify what a control holds from a screenshot, not from that call.
  • Click buttons with PostMessage, not SendMessage. A synchronous click blocks the harness for as long as the handler runs, and forever if the handler opens a modal dialog.
  • Screenshot with Graphics.CopyFromScreen, not Control.DrawToBitmap. DrawToBitmap paints children in Controls order, which is front-to-back, so a Dock = Fill control added after its docked edges is painted last and covers them. It rendered CommandReferenceForm with no filter row and no button strip — a window that is perfectly fine on screen. A harness that lies in this direction is worse than no harness.
  • Prefer probing at the Core level. A console harness against Vxi11Client reports a failure as an exception with a real message, instead of a dialog nobody can see. That is how the 64 KB truncation above was finally pinned down.
  • Never truncate the pipeline of a process talking to an instrument (... | Select-Object -First n in PowerShell kills it), because the process dies mid-transfer and leaves the instrument's output queue poisoned.
  • Form.Close() reports CloseReason.None, not UserClosing. A detached window only hands its console back on UserClosing, so programmatic closes (the Re-attach button, app shutdown) don't fight the teardown. To exercise the window's X button from a test, send WM_SYSCOMMAND / SC_CLOSE.
  • PowerShell does not wait for a GUI process launched with & unless you pipe its output (& app.exe | Out-Null), so a harness's results get read before it has written them.
  • The app holds LabEquipmentController.Core.dll while it runs, so a build fails with MSB3021/MSB3027 naming the locking process. That is a running app, not a code error; Core and Tests still build, which is why a green test run can sit next to a failed solution build.
  • A label does not tell you when its text does not fit. It clips nothing and scrolls nothing — the line is simply not drawn. A panel sized for three lines of guide text read "No local copy — expected:" and then named nothing, for every guide whose title wrapped. Nothing about that is visible from the code, and a screenshot showed it immediately.
  • sed -i rewrites a file with LF endings. In a CRLF repo with core.autocrlf=true git normalises on commit, so the diff looks clean while the working tree is left mixed — the same trap as writing LF into a CRLF JSON catalog. Re-check out anything sed touched.

Beyond the harness, two failure modes this codebase keeps meeting are worth stating once:

  • Scaling arithmetic fails by drawing, not by throwing. A waveform decoder given the wrong formula returns a plausible trace at the wrong offset. Rigol and Keysight share a preamble layout and not a formula, which makes reusing one decoder for both an easy and invisible mistake; there is a test whose only job is to fail if they ever agree.
  • A settings write that succeeds is silent. A window that saves a whole settings object built from its own controls silently reverts every field it has no control for. That cost the AI connection and its key on every app close, and nothing reported it. Load, change what you own, save.

17. Open items

Everything specified above is implemented. Confirmed against the bench: discovery, connect, identity, per-family profiles, three simultaneous console tabs, read-only queries, screen capture (1.15 MB BMP), waveform capture (1400 samples), detach into a window, re-attach by closing it, and disconnecting each console back down to none.

585 of the 11,853 entries then catalogued answered on the bench in that run, from a sweep of every query that can be sent without changing an instrument's state — 388 of the Rigol scope's, 104 of the Siglent generator's, 82 of the multimeter's. What each refused is coherent: the scope's misses are CAN triggering and bus decode, options a base DS2202 does not carry, and the meter's are the two scanner-card commands that need an -SC model. The suite that does this is Tests/Bench, off unless asked for.

Implemented but never run against the instrument it is for:

  • Four of the five waveform dialects (§11) and the Tektronix screen dump. Only the Rigol path has met hardware. The decoders are checked against each vendor's own published worked example, which is a different claim from working.
  • Thirty-three of the thirty-six catalogs (§10). Transcribed and cross-checked, not proven. Three catalogs have hardware here; there is no such instrument for the rest.

AI datasheet extraction (§11b) has run end to end against Gemini on the Siglent SDM guide, returning 211 commands where the catalog transcribed from the same document by hand holds 206, with 89% recognised. Every one of the unrecognised was checked against the guide text and appears in it verbatim: nothing was invented. That is a measurement on a guide whose answer was already known, which is what makes it worth anything — it says the pipeline works, not that an extraction from an unknown guide can be trusted without the review step.

That distinction is kept deliberately: this file is what the app is specified to do, and a specification that quietly implies bench-proof it does not have is the same failure mode as a catalog that quietly implies a guide it does not have.

Deliberately not built:

  • Dragging a console tab out of the strip to detach it — the console's Detach button and the tab's context menu cover the need.

  • Remembering open tabs across runs — only worth doing alongside a way to re-find an instrument whose DHCP address has changed (§12), otherwise launch-time reconnects are guesswork against the one-session rule (§13).

  • USB/USBTMC and GPIB (§1). Nothing is planned. What follows is the reasoning, written down because "no USB, GPIB or serial" read as one decision and was actually three — and the third of them has since been built, which is the best evidence that keeping them apart was worth the trouble.

    GPIB — no, and not for want of effort. It needs a controller card or adapter, and every one of them ships its own proprietary driver. There is no protocol to implement here that would get round that, so it would mean the NI-VISA or vendor-SDK dependency §1 exists to refuse. The instruments are also a generation older than anything on this bench, which is the smaller half of the reason.

    USB/USBTMC — no, on cost. The class protocol itself is tractable (bulk endpoints, bTag sequencing, a header per transfer), but reaching it means WinUSB on Windows and libusb elsewhere.

    Serial having landed with a native shim of its own, the distinction that matters here is worth stating precisely rather than by the word "native". System.IO.Ports does carry per-platform native code — a 15 KB libSystem.IO.Ports.Native for each non-Windows runtime — but it is Microsoft's own, restored by NuGet, and travels inside the package with nothing to install and no driver to bind. libusb is a system library the user has to already have, and on Windows a USB device speaks to WinUSB only once someone has bound that driver to it by hand. One is a build artefact; the other is a prerequisite on somebody else's machine. That is the line, and USBTMC is on the far side of it: it would either break Core's "plain net10.0, nothing Windows-only" rule or ship as a Windows-only transport the other two front ends could not offer — and a transport that exists in one build and not the others is the thing UI-SPEC §0 spends its first page arguing against.

Built after being deliberated over here (kept, because what a decision cost is worth having next to it):

  • Serial (RS-232) — was "the one that would make sense, if any did", and now exists (§13). The prediction was that SerialInstrumentClient : IInstrumentClient would be a short file because SCPI over serial is line-based exactly as ScpiClient already is, and that everything above the transport would not know the difference. Both held. What it actually cost was the framing being lifted out of ScpiClient into ScpiFraming so the two clients could share it, which the socket was better for anyway.

    The two objections held too, and became the shape of what was built rather than reasons not to:

    • Discovery did not transfer, and the shape of what replaced it is the argument. §4 is built on sweeping a subnet and asking *IDN?, and a serial port cannot be swept that way — you can enumerate the ports, but not the baud rate, parity or handshaking each instrument was set to, and probing a port that turns out to be a modem or a printer is not the harmless connect a TCP probe is.

      That first became "ports are listed and never probed, and the choosing is the user's", which was right about the danger and wrong about where the line sat. It put the whole of *IDN? behind the user typing an address, when the part that was actually unacceptable was doing any of it unasked. The line that holds is automatic: a card arriving on Serial opens nothing, and no combination of settings is ever tried behind anyone's back. A Scan somebody pressed, over the ports they chose, at the baud rates they named, with what it does written on the button, is an act they authorised — and it is the only thing that opens a port. Everything the original objection was protecting is still protected; what changed is that the protection is a button rather than an absence (§4).

      Two things the subnet sweep gets for free had to be built by hand. A port that answers nothing stays in the list, because it is still a port and still connectable at settings the scan did not try. And a reply is checked before it is believed: TCP delivers the sender's bytes or nothing, a UART at the wrong rate delivers different ones, so an "identity" is only an identity if it is printable and comma-separated.

    • It added a dependency, taking Core from one to two (§2). Costed rather than waved through: 10.5 KB gzipped in the browser payload, against PdfPig's 1.83 MB.

    Addressing was the part already done, exactly as this entry predicted: InstrumentAddress was the single place serial://COM3?baud=9600 needed reading, and all three front ends took it from there.